5.2 BIN Measurement Function (Model 3504-50, 3504-60 only)
80
9.
Press
to enter the D upper limit value setting mode.
10.
Enter the D upper limit value using the arrow keys or the numeric
keypad.
Settable Range: OFF, -199999 to 199999 (When setting to "OFF"( p. 87))
Move to the digit ........
and
Change the number ..
and
Setting a value lower than the lower limit value will not produce an error, but
accurate judgment will not be possible.
At time of shipment, it is set to ” - - - - - - ” (OFF).
11.
Press
to confirm the D upper limit value.
Enter the D lower limit value setting mode.
12.
Enter the D lower limit value using the arrow keys or the numeric
keypad.
At time of shipment, it is set to ” - - - - - - ” (OFF).
13.
Press
to confirm the D lower limit value.
(SUB display area)
Light up
Flash
Upper limit value
(SUB display area)
Flash
Lower limit value
Light up
Содержание 3504-40 C HiTester
Страница 2: ......
Страница 22: ...1 4 Names and Functions of Parts 16 ...
Страница 28: ...2 5 Turning the Power On and Off 22 ...
Страница 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Страница 66: ...4 4 Self Calibration 60 ...
Страница 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Страница 136: ...7 4 About Measurement Times 130 ...
Страница 279: ...Appendix 5 External View Appendix A7 Appendix 5 External View Unit mm 220 3 260 1 100 1 32 5 25 21 54 20 5 95 5 15 ...
Страница 286: ...Appendix 7 Initial Settings Table A14 ...
Страница 289: ......
Страница 290: ......
Страница 291: ......
Страница 292: ......