3.3 Setting the Measurement Conditions
28
Set the measurement speed.
Press
to change the mode.
Measurement speed:
FAST
,
NORM
,
SLOW
The selected item is indicated by the lit LED lamp.
The lower the measurement speed, the higher the measurement
accuracy becomes.
Measurement speed
(Allowance: ±5%±0.5 ms)
3.3.4 Measurement Speed
FAST
Measures at high speed.
NORM
Measures at normal speed.
SLOW
Measures at low speed, but provides improved measurement
accuracy.
Measurement frequency
FAST
NORM
SLOW
120 Hz
10.0 ms
37.5 ms
146.0 ms
1 kHz
2.0 ms
5.5 ms
29.5 ms
The measurement time varies depending on such factors as the open circuit,
short circuit, and load compensation ON/OFF, self calibration and the compar-
ator ON/ OFF.