6.3 Evaluate Contact Check Function
97
9
6
6.
Press
to confirm the limit value.
“LEV.ChK” (Measurement level monitoring function setting screen) will be dis-
played in the MAIN display area.
If
is not pressed the Low C reject function settings will not be saved.
7.
Press .
The unit returns to the measurement mode before the menu contents were dis-
played.
Measurement
results display
Shows Low C contact error on the right side of the D measurement value in the
MAIN display.
Output
Measure-
ment results
• The <measurement status> of the response data to the "MEASure?" measure
value command outputs “5”.
See
"Query of Measurement Data" (p. 226)
• Outputs to BIT7 of the event status register ESR1.
See
"Setting and Query of Event Status Enable Register 1 (ESER1)" (p. 215)
• Outputs Low C contact error from EXT I/O.
See
7.1 "About the EXT I/O Connector" (p. 123)