8.9 Message Reference
208
Example
Command
:CORRection:LOAD:DATA:FORMat COEFficient
:CORRection:LOAD:DATA 1.02453,0.16346
When the forwarding format setting is
COEFFICIENT
, set the load
compensation value for current measurement settings to
1.02453,0.16346
(Z_COEF,PH_COEF).
Query
:CORRection:LOAD:DATA?
:CORRection:LOAD:DATA 1.00,0.00
(when HEADER ON)
1.00,0.00
(when HEADER OFF)
When the forwarding format setting is
COEFFICIENT
, the load
compensation value for current measurement settings is set to
1.02453,0.16346
(Z_COEF,PH_COEF).
Setting and Query of Load Compensation Values