8.9 Message Reference
193
8
Setting and Query of Reference Value and Upper Limit and Lower Limit Val-
ues of First Parameter for Comparator Function in Deviation Percent Mode
Syntax
Command
Query
Response
:COMParator:FLIMit:DEViation
<Reference Value>
,
<Lower Limit
Value>,<Upper Limit Value>
:COMParator:FLIMit:DEViation?
<Reference Value>,<Lower Limit Value>,<Upper Limit Value>
<Reference Value>
= -199999 to 999999
(excluding 0)
(NR1)
<Lower Limit Value>
= OFF/ Numerical Value from -999.99 to 999.99(NR2)
<Upper Limit Value>
= OFF/ Numerical Value from -999.99 to 999.99 (NR2)
Explanation
Command Sets the reference value and upper limit and lower limit values of the
first parameter in deviation percent mode.
A numeric value in NRf format is accepted but non significant digits
are rounded off so the numeric.
Query
Returns the reference value and upper limit and lower limit value set-
tings for the first parameter in deviation percent mode in order of ref-
erence value, lower limit value, and upper limit value.
Example
Command
:COMParator:FLIMit:DEViation 250000,
-5.0,5.0
Sets 250000 for the reference value, -5% for the lower limit value, and
5% for the upper limit value of the first parameter in deviation percent
mode.
Query
Response
:COMParator:FLIMit:DEViation?
:COMPARATOR:FLIMit:DEVIATION 250000,
-5.0000,5.0000
(when HEADER ON)
250000,-5.0000,5.0000
(when HEADER OFF)
250000 is set for the reference value, -5% is set for the lower limit
value, and 5% is set for the upper limit value of the first parameter in
deviation percent mode.
Содержание 3504-40 C HiTester
Страница 2: ......
Страница 22: ...1 4 Names and Functions of Parts 16 ...
Страница 28: ...2 5 Turning the Power On and Off 22 ...
Страница 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Страница 66: ...4 4 Self Calibration 60 ...
Страница 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Страница 136: ...7 4 About Measurement Times 130 ...
Страница 279: ...Appendix 5 External View Appendix A7 Appendix 5 External View Unit mm 220 3 260 1 100 1 32 5 25 21 54 20 5 95 5 15 ...
Страница 286: ...Appendix 7 Initial Settings Table A14 ...
Страница 289: ......
Страница 290: ......
Страница 291: ......
Страница 292: ......