6.3 Evaluate Contact Check Function
100
Function to detect connection problems between the respective terminals (H
CUR
,
H
POT
, L
POT
, L
CUR
) and the object to be tested in all 4 terminal measurements.
1.
Press .
The upper part of the MAIN display area displays the menu contents and the
lower part displays the setting information.
(Refer to "Menu display organization" (p. 13) for menu order)
2.
Use
or
to select the “Cont” menu item.
3.
Use
and
to select a setting item.
Pressing
or
causes the display to change in the following manner.
6.3.3 Contact Check Function Settings
(Model 3504-60 only)
(Contact check function setting screen)
Menu content
Contact check will flash
(set in step 3)
(MAIN display area)
“
both
”
Carry out contact check before and after measuring the
test object.
“
oFF
”
Disable contact check function.
“
AFtEr
”
Carry out contact check after measuring the test object.
“
bEForE
”
Carry out contact check before measuring the test object.
Содержание 3504-40 C HiTester
Страница 2: ......
Страница 22: ...1 4 Names and Functions of Parts 16 ...
Страница 28: ...2 5 Turning the Power On and Off 22 ...
Страница 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Страница 66: ...4 4 Self Calibration 60 ...
Страница 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Страница 136: ...7 4 About Measurement Times 130 ...
Страница 279: ...Appendix 5 External View Appendix A7 Appendix 5 External View Unit mm 220 3 260 1 100 1 32 5 25 21 54 20 5 95 5 15 ...
Страница 286: ...Appendix 7 Initial Settings Table A14 ...
Страница 289: ......
Страница 290: ......
Страница 291: ......
Страница 292: ......