4.2 Load Compensation
51
4
Setting Reference Value ________________________________________
1.
In normal measurement mode, press
.
The state becomes as follows.
2.
Use
or
to select setting the load circuit compensation
menu item.
Pressing
or
causes the display to change in the following manner.
3.
Select "SEt" and press
.
(Flash)
(MAIN display area)
Flash
“AdJuSt”
Delete the disabled load circuit compensation values,
acquire the load circuit compensation value at the cur-
rently set measurement frequency and enable the load
circuit compensation value.
“SEt”
Set the standard value at the currently set measure-
ment frequency.
“oFF”
Disable load circuit compensation value at the enabled
frequency.
“on”
Enable all load circuit compensation value at all dis-
abled frequencies.
(SUB display area)
Flash
Reference value of C
Reference value of D
If you want to exit the reference value input screen and return to load compen-
sation setting screen, press
.
Содержание 3504-40 C HiTester
Страница 2: ......
Страница 22: ...1 4 Names and Functions of Parts 16 ...
Страница 28: ...2 5 Turning the Power On and Off 22 ...
Страница 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Страница 66: ...4 4 Self Calibration 60 ...
Страница 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Страница 136: ...7 4 About Measurement Times 130 ...
Страница 279: ...Appendix 5 External View Appendix A7 Appendix 5 External View Unit mm 220 3 260 1 100 1 32 5 25 21 54 20 5 95 5 15 ...
Страница 286: ...Appendix 7 Initial Settings Table A14 ...
Страница 289: ......
Страница 290: ......
Страница 291: ......
Страница 292: ......