4.3 Offset Compensation
55
4
3.
Press
to enable or disable offset compensation.
The display switches between “on” and “oFF” each time
is pressed.
4.
Press
to confirm offset compensation as enabled or dis-
abled.
When “on” is selected, the state becomes as follow.
The following will be displayed on the upper portion of the MAIN display when
“oFF” is selected.
• 3504-40, 3504-50......“Low C“ (Low C reject function setting screen)
• 3504-60.....................“Cont“ (Contact check function setting screen)
5.
Input the mantissa part of the C offset compensation value using
the arrow keys or the numeric keypad.
(If you enter a number, each digit moves one place to the right.)
Settable Range (Mantissa part)
:
±
(0.0000 to 9999.9)
Move to the digit.........
and
Change the number ...
and
(MAIN display area)
Light up
(SUB display area)
Flash
Offset compensation value mantissa part
Offset compensation value index
The smallest resolution of the C offset compensation value is 1.0E-18. If you
attempt to set it to a smaller value, the value will be set to "0".
Furthermore, if you attempt to make a setting lower than the lowest setting value
it will be set at the lowest setting value, and if you attempt to make a setting
higher than the highest setting value it will be set at the highest setting value.
Содержание 3504-40 C HiTester
Страница 2: ......
Страница 22: ...1 4 Names and Functions of Parts 16 ...
Страница 28: ...2 5 Turning the Power On and Off 22 ...
Страница 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Страница 66: ...4 4 Self Calibration 60 ...
Страница 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Страница 136: ...7 4 About Measurement Times 130 ...
Страница 279: ...Appendix 5 External View Appendix A7 Appendix 5 External View Unit mm 220 3 260 1 100 1 32 5 25 21 54 20 5 95 5 15 ...
Страница 286: ...Appendix 7 Initial Settings Table A14 ...
Страница 289: ......
Страница 290: ......
Страница 291: ......
Страница 292: ......