8.9 Message Reference
229
8
EXT I/O output
Prior-
ity
rank
Measurement
abnormality
ERR
(pin 14)
BIN1 to
BIN13
OUT OF BINS
(pin 37)
D-NG
(pin 39)
Comparator Output
(Output pin)
Panel display
Time out
LOW
HI
LOW
HI
C-HI (pin 30)
D-HI (pin 7)
H, L side contact
check errors
(after measurement)
LOW
HI
LOW
HI
C-HI (pin 30)
D-HI (pin 7)
L side contact check
error (after measure-
ment)
LOW
HI
LOW
HI
C-HI (pin 30)
D-HI (pin 7)
H side contact check
error (after measure-
ment)
LOW
HI
LOW
HI
C-HI (pin 30)
D-HI (pin 7)
H, L side contact
check errors
(before measure-
ment)
LOW
HI
LOW
HI
C-HI (pin 30)
D-HI (pin 7)
L side contact check
error (before mea-
surement)
LOW
HI
LOW
HI
C-HI (pin 30)
D-HI (pin 7)
H side contact check
error (before mea-
surement)
LOW
HI
LOW
HI
C-HI (pin 30)
D-HI (pin 7)
Fixed voltage abnor-
mality (CV error)
LOW
HI
LOW
HI
C-HI (pin 30)
D-HI (pin 7)
Over range
HI
HI
LOW
HI
C-HI (pin 30)
D-HI (pin 7)
Under range
HI
HI
LOW
HI
C-LOW (pin 31)
D-LOW (pin 8)
Outside of Low C
reject limit range
LOW LOW/ HI
LOW / HI
LOW/ HI
Standard judgment
Abnormal level
detected
LOW
HI
LOW
HI
C-HI (pin 30)
D-HI (pin 7)
Outside of C display
range
(normal measure-
ment + comparator
and BIN measure-
ment
(count setting)) *1
HI
HI
LOW
HI
C: C-HI (pin 30)
D: Standard judgment
Outside of C display
range
(comparator and BIN
measurement
(
% setting))
HI
HI
LOW
HI
C: C-HI (pin 30)
or
C-LOW (pin 31)
D: Standard judgment
Outside of D display
range
HI
HI
LOW/ HI
LOW/ HI
C: Standard judgment
D: D-HI (pin 7)
or
D-LOW (pin 8)
Outside of guaran-
teed accuracy range
HI
LOW/ HI
LOW/ HI
LOW/ HI
Standard judgment
No measurements
are made when the
power supply is
turned on
HI
HI
HI
HI
No output
Low
High
CV light will go off
Low C lights up
or
or
Содержание 3504-40 C HiTester
Страница 2: ......
Страница 22: ...1 4 Names and Functions of Parts 16 ...
Страница 28: ...2 5 Turning the Power On and Off 22 ...
Страница 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Страница 66: ...4 4 Self Calibration 60 ...
Страница 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Страница 136: ...7 4 About Measurement Times 130 ...
Страница 279: ...Appendix 5 External View Appendix A7 Appendix 5 External View Unit mm 220 3 260 1 100 1 32 5 25 21 54 20 5 95 5 15 ...
Страница 286: ...Appendix 7 Initial Settings Table A14 ...
Страница 289: ......
Страница 290: ......
Страница 291: ......
Страница 292: ......