9.1 Basic Specifications
256
Self Calibration
Decrease the drift of the measurement value
OFF, MANUAL, AUTO
In MANUAL setting, when there is a request to start from EXT I/O during
the TRG signal activation, self calibration value will be acquired and ad-
justed.
When there is a request to start from communication, self calibration will
be acquired and adjusted in the measurement after the request for self
calibration.
Average
1 to 256 (necessary)
Triggerdelay
0 to 9.999s (0.001s resolution)
Contact Check Function
(only for Model 3504-60)
Result output during contact check abnormality
EXT I/O output ...error signal output (same signal line as CV error
signal)
LED display........error shown in MAIN display
See (p. 14)
Low C Reject Function
Detect contact abnormality (detects OPEN state of 2-terminal connec-
tion).
Output as a contact error if the measurement value is lower than the judg-
ment standard.
Judgement standard ..Can be set to 0.001% to 10.000% (0.001% resolu-
tion) of the full case of the range.
Error output ................LED lights up, as well as error output from EXT I/O.
Measurement Level
Monitoring Function
Detect contact abnormalities (detect chattering)
Takes the first waveform effective value (the first half of a waveform dur-
ing FAST setting, otherwise the first waveform effective value) as a basis
for comparison, and outputs a contact error if the fluctuation of the follow-
ing waveform effective values exceeds the judgment standard.
Judgment standard ....0.01% to 100.00% (0.01% resolution) of the refer-
ence value
Error output ................
output in the main display area, as
well as error output from EXT I/O
Trigger Synchronous Output
Function
Applying a measurement signal only during measurement is possible
Key Lock Function
The setting and canceling of this function by pressing a key on the front
panel is possible
BIN Measurement (Only for
Model 3504-50, 3504-60)
The setting of 14 categories for C and upper limit and lower limit values for
D is possible (absolute value setting,
% setting)
Comparator
The setting of the upper limit and lower limit values of each of C and D is
possible (absolute value setting,
% setting)
Product Specifications
Threshold value for contact check abnormality
Range
1 to 8
Range
9, 10
Hc terminal connection resi
Hp terminal connection resistance
Approx.
500
Approx.
100
Lc terminal connection resi
Lp terminal connection resistance
Approx.
500
Approx.
100
Содержание 3504-40 C HiTester
Страница 2: ......
Страница 22: ...1 4 Names and Functions of Parts 16 ...
Страница 28: ...2 5 Turning the Power On and Off 22 ...
Страница 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Страница 66: ...4 4 Self Calibration 60 ...
Страница 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Страница 136: ...7 4 About Measurement Times 130 ...
Страница 279: ...Appendix 5 External View Appendix A7 Appendix 5 External View Unit mm 220 3 260 1 100 1 32 5 25 21 54 20 5 95 5 15 ...
Страница 286: ...Appendix 7 Initial Settings Table A14 ...
Страница 289: ......
Страница 290: ......
Страница 291: ......
Страница 292: ......