5.2 BIN Measurement Function (Model 3504-50, 3504-60 only)
75
9
5
This function enables you to set up to 14 categories of upper limit and lower limit
values for C and
one category of upper limit and lower limit values for D, and
indicates the judgment results in the BIN judgment result display area.
The corresponding signal is also output from the EXT I/O connector on the rear
of the unit.
There are two judgment modes for BIN measurement: the count value setting
and deviation percent (
%) setting
•
Count Value Setting
Set count values for the upper limit and lower limit values of the measurement
parameter.
•
Deviation Percent (
%) Setting
Enter reference values and then set percentages corresponding to the refer-
ence values for the upper limit and lower limit values.
Select judgment mode and set judgment conditions.
• Judgment mode settings ( p. 76)
• Judgment conditions settings ( p. 77)
Count Value Setting ( p. 78)
Deviation Percent (
%) Setting ( p. 82)
5.2 BIN Measurement Function
(Model 3504-50, 3504-60 only)
Measurement results display
Judgment mode
Measurement Results
Count
(Count Value Setting)
Displays the measurement value.
d-PEr (Deviation Percent
(
%) Setting)
The first parameter (C) displays the result of the
calculation:
× 100
The second parameter (D) displays the result of
the calculation: (measurement value - reference
value)
(measurement value - reference value)
| reference value|
• If the power is turned off while the unit is in BIN measurement mode, the unit
will be in BIN measurement mode when the unit is turned back on again.
• Set the upper limit and lower limit values of any BIN number that does not
require a BIN judgment to be made to OFF so that judgment will not be per-
formed.
• The measurement conditions for normal measurement mode are inherited
as is for the measurement conditions when the BIN is executed. However,
the AUTO range is automatically set to the HOLD range.
Содержание 3504-40 C HiTester
Страница 2: ......
Страница 22: ...1 4 Names and Functions of Parts 16 ...
Страница 28: ...2 5 Turning the Power On and Off 22 ...
Страница 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Страница 66: ...4 4 Self Calibration 60 ...
Страница 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Страница 136: ...7 4 About Measurement Times 130 ...
Страница 279: ...Appendix 5 External View Appendix A7 Appendix 5 External View Unit mm 220 3 260 1 100 1 32 5 25 21 54 20 5 95 5 15 ...
Страница 286: ...Appendix 7 Initial Settings Table A14 ...
Страница 289: ......
Страница 290: ......
Страница 291: ......
Страница 292: ......