8.9 Message Reference
220
Setting and Query of Header for Response Messages
Syntax
Command
Query
Response
:HEADer
<ON/ OFF>
:HEADer?
<ON/ OFF>
Explanation
Command Sets whether there is a header for response messages.
This is initialized to OFF when the power is turned on.
Query
Returns ON or OFF for the header setting of response messages.
Example
Command
:HEADer ON
Sets a header to be added to response messages.
Query
Response
:HEADer?
:HEADER ON
(when HEADER ON)
OFF
(when HEADER OFF)
A header is set to be added to response messages.
Setting and Query of Delay time for Judgement Result Output and EOM
Output Period in EXT I/O
Syntax
Command
Query
Response
:IO:OUTPut:DELay
<Numeric Value>
:IO:OUTPut:DELay?
<Numeric Value> =
0 to 0.9999 (NR1)
Explanation
Command Sets delay time comparator and the BIN judgement result
EOM
period.
A numeric value in NRf format is accepted but non significant digits
are rounded off so the numeric.
Query
Returns the setting of delay time comparator and the BIN judgement
result
EOM period as charaters.
Example
Command
:IO:OUTPut:DELay 0.0005
Sets delay time comparator and the BIN judgement result
EOM
period to 500
s.
Query
Response
:IO:OUTPut:DELay?
:IO:OUTPUT:DELay
0.0005
(when HEADER ON)
0.0005
(when HEADER OFF)
Delay time comparator and the BIN judgement result
EOM period
set at 500
s.
Note
There is an approximate error of 160
s in the delay time entered for
comparator and the BIN judgement result
EOM period for the set-
ting value.
As the delay time will vary greatly when activating the trigger from EXT
I/O during measurement, or when communicating during interface, as
far as possible, try not to control externally during measurement.
Содержание 3504-40 C HiTester
Страница 2: ......
Страница 22: ...1 4 Names and Functions of Parts 16 ...
Страница 28: ...2 5 Turning the Power On and Off 22 ...
Страница 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Страница 66: ...4 4 Self Calibration 60 ...
Страница 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Страница 136: ...7 4 About Measurement Times 130 ...
Страница 279: ...Appendix 5 External View Appendix A7 Appendix 5 External View Unit mm 220 3 260 1 100 1 32 5 25 21 54 20 5 95 5 15 ...
Страница 286: ...Appendix 7 Initial Settings Table A14 ...
Страница 289: ......
Страница 290: ......
Страница 291: ......
Страница 292: ......