5.1 Comparator Function
63
9
5
Setting the Upper Limit and Lower Limit Values for the ComparatorSet-
ting Workflow
1
Normal measurement
mode
Switch to comparator measurement mode.
The range is automatically set to HOLD.
2
Comparator measurement
mode
Switch to the setting mode for the upper limit
and lower limit values.
3
C setting
Confirm C.
Move to the next setting.
4
Reference value setting for C
(Only for the
% setting)
Set the reference value for C.
5
Upper limit and lower limit
value settings for C
Set the upper limit value for C.
Confirm the upper limit value.
Set the lower limit value for C.
Confirm the lower limit value.
6
D setting
Select D.
Confirm D.
7
Reference value setting for D
(Only for the
% setting)
Set the reference value for D.
Move to the next setting.
8
Upper limit and lower limit
value settings for D
Set the upper limit value for D.
Confirm the upper limit value.
Set the lower limit value for D.
Confirm the lower limit value.
9
Comparator measurement
mode
Switch to comparator measurement mode.
• Numeric
keypad
• Arrow keys
• Numeric
keypad
• Arrow keys
• Numeric
keypad
• Arrow keys
• Numeric
keypad
• Arrow keys
• Numeric
keypad
• Arrow keys
• Numeric
keypad
• Arrow keys
Содержание 3504-40 C HiTester
Страница 2: ......
Страница 22: ...1 4 Names and Functions of Parts 16 ...
Страница 28: ...2 5 Turning the Power On and Off 22 ...
Страница 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Страница 66: ...4 4 Self Calibration 60 ...
Страница 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Страница 136: ...7 4 About Measurement Times 130 ...
Страница 279: ...Appendix 5 External View Appendix A7 Appendix 5 External View Unit mm 220 3 260 1 100 1 32 5 25 21 54 20 5 95 5 15 ...
Страница 286: ...Appendix 7 Initial Settings Table A14 ...
Страница 289: ......
Страница 290: ......
Страница 291: ......
Страница 292: ......