8.9 Message Reference
187
8
Setting and Query of Reference Value of Second Parameter for BIN Function
in Deviation Percent Mode (Only for Models 3504-50, 3504-60)
Syntax
Command
Query
Response
:BIN:SLIMit:REFerence
<Reference Value>
:BIN:SLIMit:REFerence?
<Reference Value> = Numeric Value from -199999 to 199999 (NR1)
Explanation
Command Sets the reference value of the second parameter in deviation percent
mode.
A numeric value in NRf format is accepted but non significant digits
are rounded off so the numeric.
Query
Returns the reference value set for the second parameter in deviation
percent mode.
Example
Command
:BIN:SLIMit:REFerence 150000
Sets 150000 for the reference value of the second parameter in devi-
ation percent mode.
Query
Response
:BIN:SLIMit:REFerence?
:BIN:SLIMit:REFERENCE 150000
(when HEADER ON)
150000
(when HEADER OFF)
150000 is set for the reference value of the second parameter in devi-
ation percent mode.
Note
The upper limit and lower limit values for count value mode and the
upper limit and lower limit values for deviation percent mode of the
unit are stored separately.
Содержание 3504-40 C HiTester
Страница 2: ......
Страница 22: ...1 4 Names and Functions of Parts 16 ...
Страница 28: ...2 5 Turning the Power On and Off 22 ...
Страница 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Страница 66: ...4 4 Self Calibration 60 ...
Страница 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Страница 136: ...7 4 About Measurement Times 130 ...
Страница 279: ...Appendix 5 External View Appendix A7 Appendix 5 External View Unit mm 220 3 260 1 100 1 32 5 25 21 54 20 5 95 5 15 ...
Страница 286: ...Appendix 7 Initial Settings Table A14 ...
Страница 289: ......
Страница 290: ......
Страница 291: ......
Страница 292: ......