3.3 Setting the Measurement Conditions
31
9
3
Select a measurement range. Automatic selection is also possible.
Press
to change the mode.
Measurement Range:
AUTO
,
HOLD
The selected item is indicated by the lit LED lamp.
Measurement range
3.3.6 Measurement Range
AUTO
(Auto range)
The optimal measurement range is selected automatically.
This is useful for the measurement of unknown samples.
However, measurement takes longer.
HOLD
(Hold range)
The measurement range is fixed, and may only be altered
manually.
Take measurements in the same range regardless of the
value of the sample. This is useful for high-speed measure-
ment.
Changing the range:
When the range is changed, the decimal point and unit in
the measurement value display area change. The range
number is displayed in the SUB display area.
Reference Values
Reference Values
Current, voltage detection waveform (A/D) input range
Auto range
(Guaranteed Accuracy Range)
Содержание 3504-40 C HiTester
Страница 2: ......
Страница 22: ...1 4 Names and Functions of Parts 16 ...
Страница 28: ...2 5 Turning the Power On and Off 22 ...
Страница 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Страница 66: ...4 4 Self Calibration 60 ...
Страница 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Страница 136: ...7 4 About Measurement Times 130 ...
Страница 279: ...Appendix 5 External View Appendix A7 Appendix 5 External View Unit mm 220 3 260 1 100 1 32 5 25 21 54 20 5 95 5 15 ...
Страница 286: ...Appendix 7 Initial Settings Table A14 ...
Страница 289: ......
Страница 290: ......
Страница 291: ......
Страница 292: ......