8.9 Message Reference
240
Setting and Query of Terminator of Response Message
Syntax
Command
Query
Response
:TRANsmit:TERMinator
<Numeric Value>
<Numeric Value> = 0 to 255 (NR1)
:TRANsmit:TERMinator?
<Numeric Value> = 0/ 1 (NR1)
Explanation
Command Sets the terminator of the response message.
A numeric value in NRf format is accepted but decimals are rounded
off so the numeric value can be handled.
• When RS-232C
CR+LF : when 0
CR
: when 1 to 255
• When GP-IB (only for Models 3504-50, 3504-60)
LF+EOI
: when 0
CR+LF+EOI : when 1 to 255
Query
Returns the setting of the terminator of the response message as
NR1 numeric data of 0 and 1.
• When RS- 232C
CR+LF : when 0
CR
: when 1
• When GP-IB (only for Models 3504-50, 3504-60)
LF+EOI
: when 0
CR+LF+EOI : when 1
Example
Command
:TRANsmit:TERMinator 0
• RS- 232C
Sets the terminator to CR+LF.
• GP-IB (only for Models 3504-50, 3504-60)
Sets the terminator to LF+EOI.
Query
Response
:TRANsmit:TERMinator?
:TRANSMIT:TERMINATOR 0
(when HEADER ON)
0
(when HEADER OFF)
: The terminator is set to CR+LF.
: The terminator is set to LF+EOI.
(only for Models 3504-50, 3504-60)
Содержание 3504-40 C HiTester
Страница 2: ......
Страница 22: ...1 4 Names and Functions of Parts 16 ...
Страница 28: ...2 5 Turning the Power On and Off 22 ...
Страница 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Страница 66: ...4 4 Self Calibration 60 ...
Страница 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Страница 136: ...7 4 About Measurement Times 130 ...
Страница 279: ...Appendix 5 External View Appendix A7 Appendix 5 External View Unit mm 220 3 260 1 100 1 32 5 25 21 54 20 5 95 5 15 ...
Страница 286: ...Appendix 7 Initial Settings Table A14 ...
Страница 289: ......
Страница 290: ......
Страница 291: ......
Страница 292: ......