6.2 Trigger Delay Setting
93
9
6
This sets the delay time between when the trigger is detected and measuring
begins. With the use of this function even if measurement is commenced imme-
diately after connecting to a sample a reliable measurement value can be
attained.
1.
In normal measurement mode, press
.
The upper part of the MAIN display area displays the menu contents and the
lower part displays the setting information.
(Refer to "Menu display organization" (p. 13) for menu order)
This setting cannot be changed in comparator mode and BIN mode.
2.
Use
or
to select the “dELAY” menu item.
3.
Press
to enable or disable the trigger delay.
It will toggle “on” and “oFF” each time
is pressed.
4.
Press
to enable or disable trigger delay.
When “oFF” is selected, “SYnC” (Trigger synchronous output function setting
screen) appears in the MAIN display area.
When “on” is selected, the leftmost LED which displays the delay time will flash
in the middle row of the SUB display area.
6.2 Trigger Delay Setting
(Trigger delay setting screen)
Menu content
“on” or “oFF” flashes
(MAIN display area)
(SUB display area / “on” setting)
The delay time is lit up.
Содержание 3504-40 C HiTester
Страница 2: ......
Страница 22: ...1 4 Names and Functions of Parts 16 ...
Страница 28: ...2 5 Turning the Power On and Off 22 ...
Страница 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Страница 66: ...4 4 Self Calibration 60 ...
Страница 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Страница 136: ...7 4 About Measurement Times 130 ...
Страница 279: ...Appendix 5 External View Appendix A7 Appendix 5 External View Unit mm 220 3 260 1 100 1 32 5 25 21 54 20 5 95 5 15 ...
Страница 286: ...Appendix 7 Initial Settings Table A14 ...
Страница 289: ......
Страница 290: ......
Страница 291: ......
Страница 292: ......