8.9 Message Reference
235
8
Setting and Query of Voltage and Current Monitor Levels Display
Syntax
Command
Query
Response
:MONItor:DISPlay
<ON/ OFF>
:MONItor:DISPlay?
<ON/ OFF>
ON
:Displays measurement signal monitor level in the SUB display
area.
OFF
:Does not display the measurement signal monitor level in the
SUB display area.
Explanation
Command Sets the measurement signal monitor level display to ON or OFF.
Query
Returns the measurement signal monitor level display as ON or OFF.
Example
Command
Query
Response
:MONItor:DISPlay ON
Set the measurement signal monitor level to be displayed.
:MONItor:DISPlay?
MONITOR:DISPLAY ON
(when HEADER ON)
ON
(when HEADER OFF)
The measurement signal monitor level is set to be displayed.
Initializing the Unit
Syntax
Command
:PRESet
Explanation
Command Resets the unit to its original initialized state.
This initialized state differs from that when resetting with the
*
RST
command.
See
Содержание 3504-40 C HiTester
Страница 2: ......
Страница 22: ...1 4 Names and Functions of Parts 16 ...
Страница 28: ...2 5 Turning the Power On and Off 22 ...
Страница 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Страница 66: ...4 4 Self Calibration 60 ...
Страница 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Страница 136: ...7 4 About Measurement Times 130 ...
Страница 279: ...Appendix 5 External View Appendix A7 Appendix 5 External View Unit mm 220 3 260 1 100 1 32 5 25 21 54 20 5 95 5 15 ...
Страница 286: ...Appendix 7 Initial Settings Table A14 ...
Страница 289: ......
Страница 290: ......
Страница 291: ......
Страница 292: ......