8.9 Message Reference
239
8
Setting and Query of Trigger Synchronous Output Function
Syntax
Command
Query
Response
:SSOurce
<ON/ OFF>
:SSOurce?
<ON/ OFF>
ON : Enables the trigger synchronous output function.
OFF : Disables the trigger synchronous output function.
Explanation
Command Enables/disables the trigger synchronous output function.
Query
Returns ON or OFF for the current setting of the trigger synchronous
output function.
Example
Command
:SSOurce ON
Enables the trigger synchronous output function.
Query
Response
:SSOurce?
:SSOURCE ON
(when HEADER ON)
ON
(when HEADER OFF)
The trigger synchronous output function is enabled.
Note
When the contact check function is set to “bEForE” or “both”, the trig-
ger simultaneous output function setting cannot be changed.
Setting and Query of Wait Time for Trigger Synchronous Output Function
Syntax
Command
Query
Response
:SSOurce:WAIT
<Wait Time 1>, <Wait Time 2>
:SSOurce:WAIT?
<Wait Time 1, 2> = 0.001 to 9.999 (NR2)
<Wait Time 1> : Sets the wait time for a measurement frequency of
120 Hz.
<Wait Time 2> : Sets the wait time for a measurement frequency of
1 kHz.
Explanation
Command Sets the wait time from after the triggered output of the measurement
signal to the start of measurement.
Query
Returns the setting of the wait time for the trigger synchronous output
function as a numeric value.
Example
Command
:SSOurce:WAIT 0.500,0.250
Sets the wait time from after the trigger to the start of measurement to
500 ms for 120 Hz and 250 ms for 1 kHz.
Query
Response
:SSOurce:WAIT?
:SSOURCE:WAIT 0.500,0.250
(when HEADER ON)
0.500, 0.250
(when HEADER OFF)
The wait time from after the trigger to the start of measurement is set
to 500 ms for 120 Hz and 250 ms for 1 kHz.
Note
• When the wait time is short, measurement errors may increase.
• As for Range 9, 10, when 120 Hz, 40ms will be added to the setting
wait time and when 1 kHz, 3 ms will be added to the wait time.
Содержание 3504-40 C HiTester
Страница 2: ......
Страница 22: ...1 4 Names and Functions of Parts 16 ...
Страница 28: ...2 5 Turning the Power On and Off 22 ...
Страница 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Страница 66: ...4 4 Self Calibration 60 ...
Страница 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Страница 136: ...7 4 About Measurement Times 130 ...
Страница 279: ...Appendix 5 External View Appendix A7 Appendix 5 External View Unit mm 220 3 260 1 100 1 32 5 25 21 54 20 5 95 5 15 ...
Страница 286: ...Appendix 7 Initial Settings Table A14 ...
Страница 289: ......
Страница 290: ......
Страница 291: ......
Страница 292: ......