8.9 Message Reference
221
8
Setting and Query of Output of Judgment Result Signal Line in EXT I/O
Syntax
Command
Query
Response
:IO:RESult:RESet
<Character>
:IO:RESult:RESet?
<Character>
= ON/ OFF
ON
: Resets the judgment results when the start-of-measurement
signal (trigger signal) is input.
OFF : Updates the measurement results when measurement ends.
Explanation
Command Sets whether to reset the judgment result signal line in EXT I/O.
Query
Returns the setting of whether to reset the judgment result signal line in
EXT I/O.
Note
The judgment result signal line indicates judgment results for C or D-
HI, C or D-IN, and C or D-LO for comparator measurement and judg-
ment results OUT-OF-BINS, D-NG, BIN1 to BIN14 for BIN measure-
ment, and error (CV error, Low C error, level detected error, contact
error).
See
7.1 "About the EXT I/O Connector" (p. 123)
Example
Command
:IO:
RESult:RESet OFF
Sets the judgment results to be updated when measurement ends.
Query
Response
:IO:
RESult:RESet
?
:IO:
RESULT:RESET OFF
(when HEADER ON)
OFF
(when HEADER OFF)
The judgment results are set to be updated when judgment ends.
Содержание 3504-40 C HiTester
Страница 2: ......
Страница 22: ...1 4 Names and Functions of Parts 16 ...
Страница 28: ...2 5 Turning the Power On and Off 22 ...
Страница 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Страница 66: ...4 4 Self Calibration 60 ...
Страница 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Страница 136: ...7 4 About Measurement Times 130 ...
Страница 279: ...Appendix 5 External View Appendix A7 Appendix 5 External View Unit mm 220 3 260 1 100 1 32 5 25 21 54 20 5 95 5 15 ...
Страница 286: ...Appendix 7 Initial Settings Table A14 ...
Страница 289: ......
Страница 290: ......
Страница 291: ......
Страница 292: ......