9.2 Accuracy
260
* Not certified guaranteed when range is No.1 and No. 2 for measurement
signal level 100 mV.
Support for Range Numbers and Range Names
9
C
±1.2%rdg.±50dgt.
±0.7%rdg.±40dgt.
±1.8%rdg.±75dgt.
±1.05%rdg.±60dgt.
D
±0.0060
±0.0050
±0.0090
±0.0075
10
C
±2.5%rdg.±50dgt.
±2.0%rdg.±40dgt.
±3.75%rdg.±75dgt.
±3.0%rdg.±60dgt.
D
±0.0200±0.008xC
H
±0.0180±0.08xC
H
±0.0300±0.012xC
H
±0.0270±0.12xC
H
C-D
(CL: electrostatic capacitance [pF] of sample, , CH: electrostatic capacitance [mF] of sample)
when D
0.1
Rang
e No.
Frequen-
cy Pa-
rameter
Guaranteed Accuracy of 6 Months
Guaranteed Accuracy of 1 Year
120 Hz
1 kHz
120 Hz
1 kHz
Yes: Available/ No: Unavailable
Parameter
C
CV
Operation
Frequency
Range No.
120 Hz
1 kHz
1
200 pF
20 pF
Yes
2
2 nF
200 pF
Yes
3
20 nF
2 nF
Yes
4
200 nF
20 nF
Yes
5
2
F
200 nF
Yes
6
20
F
2
F
Yes
7
200
F
20
F
Yes
8
0.7 mF (when 1 V)
1.45 mF (when 500 mV, 100
mV)
70
F (when 1 V)
170
F (when 500 mV, 100
mV)
Yes
9
2 mF
200
F
No
10
20 mF
2 mF
No
When D>0.1, the measurement values are the reference values.
Содержание 3504-40 C HiTester
Страница 2: ......
Страница 22: ...1 4 Names and Functions of Parts 16 ...
Страница 28: ...2 5 Turning the Power On and Off 22 ...
Страница 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Страница 66: ...4 4 Self Calibration 60 ...
Страница 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Страница 136: ...7 4 About Measurement Times 130 ...
Страница 279: ...Appendix 5 External View Appendix A7 Appendix 5 External View Unit mm 220 3 260 1 100 1 32 5 25 21 54 20 5 95 5 15 ...
Страница 286: ...Appendix 7 Initial Settings Table A14 ...
Страница 289: ......
Страница 290: ......
Страница 291: ......
Страница 292: ......