5.2 BIN Measurement Function (Model 3504-50, 3504-60 only)
86
17.
Press
to confirm the D upper limit value.
Enter the D lower limit value setting mode.
18.
In the same way enter the D lower limit value using the arrow keys
or the numeric keypad.
At time of shipment, it is set to ” - - - - - - ” (OFF).
19.
Press
to confirm the D lower limit value.
20.
Press .
Enter the BIN measurement mode.
See
"Performing BIN Measurement" (p. 88)
(SUB display area)
Light up
Flash
Lower limit value
(Light up)
• The reference value for the
setting and
% setting become display count
values that are independent of the measurement conditions. If the measure-
ment conditions differ, the absolute values that signify the count values
change.
For example, if the range is currently set at 100 pF this would mean that C
count value of 50000 = 50E-9 and D count value 100 = 0.00100.
• The measurement conditions for normal measurement mode are used for
the BIN measurement mode. Set the measurement conditions to use for
BIN measurement mode while the unit is in normal measurement mode.
• When checking the upper/lower limit values, error judgment is not done.
• Check the following because judgment cannot be performed properly if the
upper limit and lower limit values are set incorrectly.
•Are the setting values within the display range of the measurement range?
•Is the large/small relationship of the upper limit and lower limit values
correct?
• After settings are complete,
can be used to toggle the judgment
range display to confirm the judgment range.
Содержание 3504-40 C HiTester
Страница 2: ......
Страница 22: ...1 4 Names and Functions of Parts 16 ...
Страница 28: ...2 5 Turning the Power On and Off 22 ...
Страница 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Страница 66: ...4 4 Self Calibration 60 ...
Страница 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Страница 136: ...7 4 About Measurement Times 130 ...
Страница 279: ...Appendix 5 External View Appendix A7 Appendix 5 External View Unit mm 220 3 260 1 100 1 32 5 25 21 54 20 5 95 5 15 ...
Страница 286: ...Appendix 7 Initial Settings Table A14 ...
Страница 289: ......
Страница 290: ......
Страница 291: ......
Страница 292: ......