6.3 Evaluate Contact Check Function
99
9
6
6.
Press
to confirm the limit value.
“JudGE”(Judgment mode setting screen) will be displayed in the MAIN display
area.
If
is not pressed, the measurement level monitoring function settings
will not be saved.
7.
Press .
The unit returns to the measurement mode it was in prior to the menu items
being displayed.
Measurement
results display
will be displayed in the MAIN display area.
See
"MAIN display area ERROR display" (p. 14)
Output
Measure-
ment results
• The response data to the “:MEASure?” measure value command is output
thusly: <measurement status> 4, <C measurement value>66, and
<D measurement value>666666.
See
"Query of Measurement Data" (p. 226)
• Outputs to BIT5 of event status register ESR0.
See
"Setting and Query of Event Status Enable Register 1 (ESER1)" (p. 215)
• Outputs detected level abnormalities from EXT I/O.
See
7.1 "About the EXT I/O Connector" (p. 123)
Even in the case of normal contact, error output may occur due to the influence
of external noise. (It is possible to judge external noise.)
Содержание 3504-40 C HiTester
Страница 2: ......
Страница 22: ...1 4 Names and Functions of Parts 16 ...
Страница 28: ...2 5 Turning the Power On and Off 22 ...
Страница 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Страница 66: ...4 4 Self Calibration 60 ...
Страница 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Страница 136: ...7 4 About Measurement Times 130 ...
Страница 279: ...Appendix 5 External View Appendix A7 Appendix 5 External View Unit mm 220 3 260 1 100 1 32 5 25 21 54 20 5 95 5 15 ...
Страница 286: ...Appendix 7 Initial Settings Table A14 ...
Страница 289: ......
Страница 290: ......
Страница 291: ......
Страница 292: ......