8.9 Message Reference
195
8
Setting and Query of Reference Value and Upper Limit and Lower Limit Val-
ues of Second Parameter for Comparator Function in Deviation Percent Mode
Syntax
Command
Query
Response
:COMParator:SLIMit:DEViation
<Reference Value>
,
<Lower
Limit Value>,<Upper Limit Value>
:COMParator:SLIMit:DEViation?
<Reference Value>,<Lower Limit Value>,<Upper Limit Value>
<Reference Value>
= -199999 to 199999(NR1)
<Lower Limit Value>
= OFF/ Numeric Value from -199999 to 199999 (NR1)
<Upper Limit Value>
= OFF/ Numeric Value from -199999 to 199999 (NR1)
Explanation
Command Sets the reference value and upper limit and lower limit values of the
second parameter in deviation percent mode.
A numeric value in NRf format is accepted but non significant digits
are rounded off so the numeric.
Query
Returns the reference value and upper limit and lower limit value set-
tings for the second parameter in deviation percent mode in order of
reference value, lower limit value, and upper limit value.
Example
Command
:COMParator:SLIMit:DEViation 2000,-5,5
Sets 2000 for the reference value, -5 for the lower limit value, and 5
for the upper limit value of the second parameter in deviation percent
mode.
Query
Response
:COMParator:SLIMit:DEViation?
:COMPARATOR:SLIMit:DEVIATION 2000,-5,5
(when HEADER ON)
2000,-5,5
(when HEADER OFF)
2000 is set for the reference value, -5 is set for the lower limit value,
and 5 is set for the upper limit value of the second parameter in devia-
tion percent mode.
Note
The measurement value for the second parameter in deviation per-
cent mode is the result of the calculation (measurement value - refer-
ence value).
Содержание 3504-40 C HiTester
Страница 2: ......
Страница 22: ...1 4 Names and Functions of Parts 16 ...
Страница 28: ...2 5 Turning the Power On and Off 22 ...
Страница 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Страница 66: ...4 4 Self Calibration 60 ...
Страница 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Страница 136: ...7 4 About Measurement Times 130 ...
Страница 279: ...Appendix 5 External View Appendix A7 Appendix 5 External View Unit mm 220 3 260 1 100 1 32 5 25 21 54 20 5 95 5 15 ...
Страница 286: ...Appendix 7 Initial Settings Table A14 ...
Страница 289: ......
Страница 290: ......
Страница 291: ......
Страница 292: ......