8.9 Message Reference
206
Setting and Query of Load Compensation Function
Syntax
Command
Query
:CORRection:LOAD
<Character>
<Character> = ON/ OFF/ RETurn
:CORRection:LOAD?
<Character> = ON/ SPOT/ OFF
Explanation
Command Sets the load compensation function.
If ON is set, load compensation data begins to be incorporated based
on the current measurement conditions (frequency, level, range,
self calibration, open circuit compensation, short circuit compensa-
tion) and reference values for load compensation conditions. After
incorporating of the data ends properly, load compensation is
enabled. If it does not end properly, the load compensation data
remains the same as last time.
ON
: Delete disabled load circuit compensation value, acquire
load circuit compensation value at the current measurement
conditions and enable the load circuit compensation function.
OFF
: Disable the load circuit compensation value at the enabled
measurement frequency.
RETurn: Enable all disabled load circuit compensation values.
Query
Returns the setting of the load compensation function as characters.
ON
: The load compensation function is enabled for the current
measurement conditions.
SPOT : The load compensation function is enabled for other than the
current measurement conditions.
OFF : The load compensation function is disabled.
Example
Command
:CORRection:LOAD ON
Delete disabled load circuit compensation value, acquire load circuit
compensation value at the current measurement conditions and
enable the load circuit compensation function.
Query
Response
:CORRection:LOAD?
:CORRECTION:LOAD ON
(when HEADER ON)
ON
(when HEADER OFF)
The load compensation function is enabled for the current measure-
ment conditions.
Содержание 3504-40 C HiTester
Страница 2: ......
Страница 22: ...1 4 Names and Functions of Parts 16 ...
Страница 28: ...2 5 Turning the Power On and Off 22 ...
Страница 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Страница 66: ...4 4 Self Calibration 60 ...
Страница 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Страница 136: ...7 4 About Measurement Times 130 ...
Страница 279: ...Appendix 5 External View Appendix A7 Appendix 5 External View Unit mm 220 3 260 1 100 1 32 5 25 21 54 20 5 95 5 15 ...
Страница 286: ...Appendix 7 Initial Settings Table A14 ...
Страница 289: ......
Страница 290: ......
Страница 291: ......
Страница 292: ......