8.9 Message Reference
185
8
Setting and Query of Upper Limit and Lower Limit Values of Second
Parameter for BIN Function in Count Value Mode
(Only for Models 3504-50, 3504-60)
Syntax
Command
Query
Response
:BIN:SLIMit:COUNt
<Lower Limit Value>,<Upper Limit Value>
:BIN:SLIMit:COUNt?
<Lower Limit Value>,<Upper Limit Value>
<Lower Limit Value> =
OFF/ Numeric Value from -199999 to 199999 (NR1)
<Upper Limit Value>
= OFF/ Numeric Value from -199999 to 199999 (NR1)
Explanation
Command Sets the upper limit and lower limit values of the second parameter in
count value mode.
A numeric value in NRf format is accepted but non significant digits
are rounded off so the numeric value can be handled.
Query
Returns the upper limit and lower limit value settings for the second
parameter in count value mode in order of lower limit value and upper
limit value.
Example
Command
:BIN:SLIMit:COUNt 100000,150000
Sets 100000 for the lower limit value and 150000 for the upper limit
value of the second parameter in count value mode.
Query
Response
:BIN:SLIMit:COUNt?
:BIN:SLIMit:COUNT 100000,150000
(when HEADER ON)
100000,150000
(when HEADER OFF)
100000 is set for the lower limit value and 150000 is set for the upper
limit value of the second parameter in count value mode.
Содержание 3504-40 C HiTester
Страница 2: ......
Страница 22: ...1 4 Names and Functions of Parts 16 ...
Страница 28: ...2 5 Turning the Power On and Off 22 ...
Страница 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Страница 66: ...4 4 Self Calibration 60 ...
Страница 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Страница 136: ...7 4 About Measurement Times 130 ...
Страница 279: ...Appendix 5 External View Appendix A7 Appendix 5 External View Unit mm 220 3 260 1 100 1 32 5 25 21 54 20 5 95 5 15 ...
Страница 286: ...Appendix 7 Initial Settings Table A14 ...
Страница 289: ......
Страница 290: ......
Страница 291: ......
Страница 292: ......