8.9 Message Reference
209
8
Setting and Query of Output Format for Load Compensation Values
Syntax
Command
Query
Response
:CORRection:LOAD:DATA:FORMat
<Character>
:CORRection:LOAD:DATA:FORMat?
<Character>
= COEFficient/ ZPH/ CD
COEFficient: The impedance compensation coefficient and phase
compensation coefficient.
ZPH : The actual measurement values for the impedance and phase.
CD
: The actual measurement values for C and D.
Explanation
Command Sets the forwarding format of load compensation values.
Query
Returns the setting of the forwarding format for load compensation
values.
Example
Command
:CORRection:LOAD:DATA:FORMat COEFficient
Sets the forwarding format of load compensation values to be the
impedance compensation coefficient and phase compensation coeffi-
cient.
Query
Response
:CORRection:LOAD:DATA:FORMat?
:CORRECTION:LOAD:DATA:FORMAT COEFFICIENT
(when
HEADER ON)
COEFFICIENT
(when HEADER OFF)
The forwarding format of load compensation values are set to be the
impedance compensation coefficient and phase compensation coeffi-
cient.
Содержание 3504-40 C HiTester
Страница 2: ......
Страница 22: ...1 4 Names and Functions of Parts 16 ...
Страница 28: ...2 5 Turning the Power On and Off 22 ...
Страница 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Страница 66: ...4 4 Self Calibration 60 ...
Страница 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Страница 136: ...7 4 About Measurement Times 130 ...
Страница 279: ...Appendix 5 External View Appendix A7 Appendix 5 External View Unit mm 220 3 260 1 100 1 32 5 25 21 54 20 5 95 5 15 ...
Страница 286: ...Appendix 7 Initial Settings Table A14 ...
Страница 289: ......
Страница 290: ......
Страница 291: ......
Страница 292: ......