8.9 Message Reference
236
Setting and Query of Measurement Range
Syntax
Command
Query
Response
:RANGe
<Numeric Value>
:RANGe?
<Numeric Value>
<Numeric Value>
= 1 to 10 (NR1)
Explanation
Command Sets the measurement range.
A numeric value in NRf format is accepted but decimals are rounded
off so the numeric value can be handled.
If this command is executed, the setting of the :RANGe:AUTO com-
mand is automatically changed to OFF.
If this command is executed when the setting of the equivalent circuit
is set to AUTO, the setting (SER/PAR) of the equivalent circuit is auto-
matically changed to the optimal setting.
Query
Returns the setting of the measurement range as an NR1 value from
1 to 10.
Example
Command
:RANGe 5
Sets the measurement range to 5 (200 nF) (when 1 kHz).
Query
Response
:RANGe?
:RANGE 5
(when HEADER ON)
5
(when HEADER OFF)
The measurement range is set to 5 (200 nF) (when 1 kHz).
Note
Numerical Value
(Range Number)
Measurement Frequency
120Hz
1kHz
1 200
pF
20
pF
2 2
nF
200
pF
3
20 nF
2 nF
4 200
nF
20
nF
5 2
F
200 nF
6
20
F
2
F
7
200
F
20
F
8
0.7 mF (when 1 V),
1.45 mF (when 500 mV, 100 mV)
70
F (when 1 V),
170
F (when 500 mV, 100 mV)
9 2
mF
200
F
10
20 mF
2 mF
Содержание 3504-40 C HiTester
Страница 2: ......
Страница 22: ...1 4 Names and Functions of Parts 16 ...
Страница 28: ...2 5 Turning the Power On and Off 22 ...
Страница 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Страница 66: ...4 4 Self Calibration 60 ...
Страница 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Страница 136: ...7 4 About Measurement Times 130 ...
Страница 279: ...Appendix 5 External View Appendix A7 Appendix 5 External View Unit mm 220 3 260 1 100 1 32 5 25 21 54 20 5 95 5 15 ...
Страница 286: ...Appendix 7 Initial Settings Table A14 ...
Страница 289: ......
Страница 290: ......
Страница 291: ......
Страница 292: ......