7.4 About Measurement Times
128
Measurement times differ depending on the measurement conditions. Refer to
the following values.
Analog Measure-
ment Signal
INDEX
The output time (
T3
) for an analog measurement signal (INDEX) can be
obtained by the following equation.
T3 = (1 or 2) + 3
(HOLD range contact check: OFF, trigger simultaneous output function: disabled)
1. No request for self calibration
(Allowable tolerance:
±
5%
±
0.3 ms)
2. Request for self calibration
3. Contact check evaluation time (Only for 3504-60)
Calculation times differ depending on the execution of the contact checks.
* When the contact check setting is at “bEForE”, “both” the analog measurement
time will be delayed because automatically after contact check, the measure-
ment will start only after the wait time for the trigger simultaneous output func-
tion has been activated.
The above values are the initial values of the wait time settings. They are for
reference only. The actual value will be 5.0 ms + wait time of the trigger simul-
taneous output function.
A wait time for trigger synchronous output function will be decided according to
the following table.
As for Range 9 and 10, when 120 Hz, 40 ms will be added to the wait time of
Range 1 to 8, when 1 KHz, 3 ms will be added to the wait time.
7.4 About Measurement Times
• All of the values are reference values. Note that they may differ depending
on the conditions of use.
• A wait of 350 ms is included when the frequency, level, and range change.
Measurement Frequency
Measurement Speed
FAST (ms)
NORM (ms)
SLOW (ms)
120 Hz
8.3
33.3
133.3
1 kHz
1
4
24
Measurement Frequency
Measurement Speed
FAST (ms)
NORM (ms)
SLOW (ms)
120 Hz
24.1
41
141
1 kHz
3.7
5.7
26.9
Measurement Frequency
Measurement Speed
BEFORE (ms)
AFTER (ms)
BOTH (ms)
120 Hz
15.0 *
5.0
20.0 *
1 kHz
7.0 *
5.0
12.0 *
Wait time default setting
Measurement
Frequency
Range
1 to 8
Range
9, 10
120 Hz
10 ms
10 ms + 40 ms
1 kHz
2 ms
2 ms + 3 ms
Содержание 3504-40 C HiTester
Страница 2: ......
Страница 22: ...1 4 Names and Functions of Parts 16 ...
Страница 28: ...2 5 Turning the Power On and Off 22 ...
Страница 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Страница 66: ...4 4 Self Calibration 60 ...
Страница 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Страница 136: ...7 4 About Measurement Times 130 ...
Страница 279: ...Appendix 5 External View Appendix A7 Appendix 5 External View Unit mm 220 3 260 1 100 1 32 5 25 21 54 20 5 95 5 15 ...
Страница 286: ...Appendix 7 Initial Settings Table A14 ...
Страница 289: ......
Страница 290: ......
Страница 291: ......
Страница 292: ......