8.9 Message Reference
192
Setting and Query of Upper Limit and Lower Limit Values of First Parame-
ter for Comparator Function in Count Value Mode
Syntax
Command
Query
Response
:COMParator:FLIMit:COUNt
<Lower Limit Value>,<Upper Limit
Value>
:COMParator:FLIMit:COUNt?
<Lower Limit Value>,<Upper Limit Value>
<Lower Limit Value>
= OFF/ Numeric Value from -199999 to 999999 (NR1)
<Upper Limit Value>
= OFF/ Numeric Value from -199999 to 999999 (NR1)
Explanation
Command Sets the upper limit and lower limit values of the first parameter for the
comparator function in count value mode.
A numeric value in NRf format is accepted but non significant digits
are rounded off so the numeric.
Query
Returns data for the upper limit and lower limit value settings of the
first parameter for the comparator function in order of lower limit value
and upper limit value.
Example
Command
:COMParator:FLIMit:COUNt 112345,123456
Sets 112345 for the lower limit value and 123456 for the upper limit
value of the first parameter in count value mode.
Query
Response
:COMParator:FLIMit:COUNt?
:COMPARATOR:FLIMIT:COUNT 112345, 123456
(when HEADER ON)
112345,123456
(when HEADER OFF)
112345 is set for the lower limit value and 123456 is set for the upper
limit value of the first parameter in count value mode.
Содержание 3504-40 C HiTester
Страница 2: ......
Страница 22: ...1 4 Names and Functions of Parts 16 ...
Страница 28: ...2 5 Turning the Power On and Off 22 ...
Страница 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Страница 66: ...4 4 Self Calibration 60 ...
Страница 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Страница 136: ...7 4 About Measurement Times 130 ...
Страница 279: ...Appendix 5 External View Appendix A7 Appendix 5 External View Unit mm 220 3 260 1 100 1 32 5 25 21 54 20 5 95 5 15 ...
Страница 286: ...Appendix 7 Initial Settings Table A14 ...
Страница 289: ......
Страница 290: ......
Страница 291: ......
Страница 292: ......