8.9 Message Reference
197
8
Setting and Query of Open Circuit Compensation Function
Syntax
Command
Query
:CORRection:OPEN
<Character>
<Character>ALL/ ON/ OFF/ RETurn
:CORRection:OPEN?
<Character>
=ALL/ ON/ SPOT/ OFF
Explanation
Command Sets the open circuit compensation function.
If ALL or ON is set, open circuit compensation value begins to be
incorporated and then open circuit compensation is enabled after
incorporation ends.
ALL
: Delete disabled open circuit compensation value, acquire
open circuit compensation value according to the measure-
ment values set at open circuit compensation reading set-
ting command (:CORRection:OPEN:POINt) and enable the
open circuit compensation function.
ON
:
Delete disabled open circuit compensation value, acquire
open circuit compensation value at the current measure-
ment conditions (frequency, level) and enable the open cir-
cuit compensation function.
OFF
: Disable open circuit compensation value at the enabled
measurement conditions (frequency, level).
RETurn: Enable all disabled open circuit compensation values.
Query
Returns the setting of the open circuit compensation function as char-
acters.
ALL : The open circuit compensation function is enabled for all mea-
surement conditions (frequency, level).
ON
: The open circuit compensation function is enabled for the cur-
rent measurement conditions (frequency, level).
SPOT : The open circuit compensation function is enabled for other
than the current measurement conditions (frequency, level).
OFF : The open circuit compensation function is disabled.
Example
Command
:CORRection:OPEN ALL
Delete disabled open circuit compensation value, acquire open circuit
compensation value according to the measurement values set at
:CORRection:OPEN:POINt and enable the open circuit compensation
function.
Query
Response
:CORRection:OPEN?
:CORRECTION:OPEN ALL
(when HEADER ON)
ALL
(when HEADER OFF)
The compensation function is enabled for all measurement conditions
(frequency, level).
Содержание 3504-40 C HiTester
Страница 2: ......
Страница 22: ...1 4 Names and Functions of Parts 16 ...
Страница 28: ...2 5 Turning the Power On and Off 22 ...
Страница 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Страница 66: ...4 4 Self Calibration 60 ...
Страница 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Страница 136: ...7 4 About Measurement Times 130 ...
Страница 279: ...Appendix 5 External View Appendix A7 Appendix 5 External View Unit mm 220 3 260 1 100 1 32 5 25 21 54 20 5 95 5 15 ...
Страница 286: ...Appendix 7 Initial Settings Table A14 ...
Страница 289: ......
Страница 290: ......
Страница 291: ......
Страница 292: ......