![YOKOGAWA ScopeCorder DL750 User Manual Download Page 92](http://html.mh-extra.com/html/yokogawa/scopecorder-dl750/scopecorder-dl750_user-manual_911347092.webp)
11-54
IM 701210-06E
File Name When Action Is Set to Save to File or Image
If the Auto Naming of the File menu or the Image Save menu is OFF, numbers are
automatically assigned to the file names. Otherwise, the file is saved using the specified
method.
For details, see section 13.11, “Saving the Screen Image Data” or section 13.7, “Saving/
Loading Waveform Data.”
Notes When Performing GO/NO-GO Determination
• The determination results (the number of determinations and failures) are displayed.
Determination count
Fail count
Measured values for each waveform parameter
XXXXXOOOO--------
Indicates whether the condition of
each waveform parameter 1 to 16 is met.
Waveform parameters
In this example, the condition of waveform
parameter 1 is not met and the condition of
waveform parameter 6 is met.
O: The specified condition is met
X: The specified condition is not met
-: Waveform parameter for which a
condition is not specified
1,2,3, . . . . . . . . . . . ,16:
waveform parameter
• All keys other than START/STOP and the Abort soft key are disabled during the
determination. (When Remote is ON, the Exec (One Shot) soft key is also valid.)
• When GO/NO-GO determination is executed, the trigger mode is automatically
changed to Single.
• Determination is not possible when the acquisition mode is set to Average.
• The determination period is synchronized to the trigger. (However, triggers are not
accepted while the action after determination is being carried out.)
• Determination is not possible when realtime recording is in progress or when
waveforms are being acquired using the dual capture function.
• If any of the following actions is activated while the DL750/DL750P is being accessed
through the FTP server function or the Web server function, the action is not executed
until the access is finished.
Printing/saving of the screen image data and saving of the waveform data
Notes When Action Is Set to Save to File or Image Save
• Do not specify the root directory of the medium as the save destination directory.
(Only up to 512 files can be saved to the root directory of a medium that has been
formatted using the DL750/DL750P.)
• The maximum number of files that can be created in a single directory is 5000. Do
not place files in the save destination folder before starting GO/NO-GO determination.
• If you select Save to File and Image Save simultaneously, separate the folders
specified by the FILE menu and the IMAGE SAVE menu.
• If you selected Numbering for the file name assignment method (Auto Naming) in the
FILE menu and the IMAGE SAVE menu, the creation of files takes an extended
period when the number of saved files becomes large. When creating more than
2000 files, select Date for Auto Naming.
11.9 GO/NO-GO Determination Using Measured Waveform Parameters