Specifications
19
Item
Specifications
Immunity
Complying standard
EN61326-1 Table 2 (for use in industrial locations)
EN61326-2-1
Already certified (701210/701230/701250/701251/701255/701260/
701261/701262/701265/701270/701271/701275/701280/700986/700987/
701955/701956/701957/701958)
• Influence in the immunity environment (performance criterion A)
Noise increase
701250:
≤
±
5 mV (1:1 input, 5 mV/div conversion)
701251:
≤
±
3 mV (1:1 input, 1 mV/div conversion)
701255:
≤
±
25 mV (1:1 input, 5 mV/div conversion)
701260:
≤
±
30 mV (1:1 input, 20 mV/div conversion)
701261/701262:
≤
±
3 mV
701265:
≤
±
0.05 mV
701270:
≤
±
100
µ
STR (when equivalent to
±
100 mV,
gauge factor = 2, and bridge voltage = 2 V)
701271:
≤
±
100
µ
STR (when equivalent to
±
100 mV,
gauge factor = 2, and bridge voltage = 2 V)
701275:
≤
±
3 mV (1:1 input, 5 mV/div conversion)
701280: Within the specifications
Test condition 701250:
10 MS/s, envelope mode, 50 mV/div, no input filter,
with the tip of the probe (700929 (10:1)) shorted
701251:
1 MS/s, envelope mode, 10 mV/div, no input filter,
with the tip of the probe (700929 (10:1)) shorted
701255:
10 MS/s, envelope mode, 50 mV/div, no input filter,
with the tip of the probe (701940 (10:1)) shorted
701260:
100 kS/s, envelope mode, 20 mV/div, no input filter,
with the tip of the probe (700929 (10:1)) shorted
701261/701262: 100 kS/s, envelope mode, 5 mV/div, no input filter,
with the end of the 3-m twisted-pair cable shorted
701265:
100 kS/s, envelope mode, 0.1 mV/div, no input filter,
With the end of the 3-m twisted-pair cable shorted
701270:
100 kS/s, envelope mode, 500
µ
STR, gauge factor: 2.0, no input
filter, 701955 bridge voltage: 2 V, 701956 bridge voltage: 10 V
701271:
100 kS/s, envelope mode, 500
µ
STR, gauge factor: 2.0, no input
filter, 701957 bridge voltage: 2 V, 701958 bridge voltage: 10 V
701275:
100 kS/s, envelope mode, 50 mV/div, no input filter,
with the tip of the probe (700929 (10:1)) shorted
701280:
25 kS/s, envelope mode, Frequency, 0.1 Hz/div, no input filter,
User (
±
1 V, threshold level: 0 V, hysteresis:
±
1%), with the tip of
the probe (700929 (10:1)) shorted
Test Item
1. Electrostatic discharge
EN61000-4-2
Air discharge:
±
8 kV, contact discharge:
±
4 kV,
criteria B
2. Radiated immunity
EN61000-4-3
80 MHz to 1 GHz, 1.4 GHz to 2 GHz, 10 V/m
Criteria A
3. Conducted immunity
EN61000-4-6
3 V, criteria A
4. Electrical fast transient/burst
EN61000-4-4
Power line:
±
2 kV, signal line:
±
1 kV, criteria B
5. Power frequency magnetic field EN61000-4-8
30 A/m, 50 Hz, criteria A
6. Surge immunity
EN61000-4-5
Between lines:
±
1 kV, common:
±
2 kV, criteria B
7. Voltage dip and interruption
EN61000-4-11
0.5 cycle, both polarities, 100%, criteria A
Definitions of criteria
• Criteria A
Aforementioned “Influence in the immunity environment” is
A and B
met during the test.
• Criteria B
This apparatus continues to operate without hang-up or
falling into uncontrollable conditions during the test. No
change of actual operating state or stored data is allowed.
1.
The Overvoltage Category (Installation Category) is a value used to define the transient overvoltage condition and includes
the impulse withstand voltage regulation. I applies to electrical equipment that is powered by a circuit with overvoltage
control. II applies to electrical equipment that is powered by a fixed installation such as a distribution board.
2.
Measurement category II (CAT II) applies to measurement of electrical equipment which is supplied from fixed installations
such as a wall outlet wired from a distribution board, or of the wires themselves.
3.
Pollution Degree: Applies to the degree of adhesion of a solid, liquid, or gas which deteriorates withstand voltage or surface
resistivity. Pollution Degree 2 applies to normal indoor atmospheres (with only non-conductive pollution).
4.
Use cables of length 3 m or less.
5.
Use cables of length 30 m or less.
19.11 General Specifications