![YOKOGAWA ScopeCorder DL750 User Manual Download Page 323](http://html.mh-extra.com/html/yokogawa/scopecorder-dl750/scopecorder-dl750_user-manual_911347323.webp)
19-35
IM 701210-06E
Specifications
19
Acceleration/Voltage Module (with AAF) (701275) Specifications
Item
Specifications
Standard operating
Temperature: 23
°
C
±
5
°
C
conditions
Humidity: 55%
±
10% RH
After a 30-minute warm-up and after calibration
Effective measurement
20 div (display range: 10 div)
range
Number of input channels
2
Input coupling settings
AC, DC, ACCL (acceleration), and GND
Maximum sample rate
100 kS/s
Input format
Isolated unbalanced
Frequency characteristics
1
(–3 dB point when a sine wave of amplitude
±
3 divisions is input)
Waveform measurement mode:
DC to 40 kHz
Acceleration measurement mode:
0.4 Hz to 40 kHz
Voltage-axis sensitivity
When using 1:1 probe attenuation:
5 mV/div to 10 V/div (1-2-5 steps)
setting
When using 10:1 probe attenuation:
50 mV/div to 100 V/div (1-2-5 steps)
(combined with the recommended probe 701940)
Acceleration (
±
5 V =
×
1 range):
×
0.1 to
×
1 to
×
100 (in 1-2-5 steps)
Maximum input voltage (at a frequency of 1 kHz or less)
42 V (DC+Acpeak)
2
Maximum allowable common mode voltage (at a frequency of 1 kHz or less)
Working voltage of safety standard
30 Vrms (CAT I and CAT II)
3
DC offset Maximum selectable range
±
5 div
Vertical (voltage) axis
Waveform measurement mode
DC accuracy:
±
(0.25% of 10 div)
accuracy
Acceleration measurement mode:
±
(0.5% of 10 div) at 1 kHz
Input connector
Metal BNC connector
Input impedance –3dB
1 M
Ω ±
1%, approx. 35 pF
point when AC coupled Low0.4 Hz or less (0.04 Hz or less when using the 701940) (typical
4
)
frequency attenuation point
Common mode rejection ratio
80 dB (50/60 Hz) or more (typical
4
)
Residual noise level
±
100
µ
V or
±
0.01 div, whichever is greater (typical
4
)
(Input section shorted)
A/D conversion resolution
16 bits (2400 LSB/div)
Temperature coefficient
When in waveform measurement mode (excluding AUTO filter)
Zero point:
±
(0.02% of 10 div)/
°
C (typical
4
)
Gain:
±
(0.02% of 10 div)/
°
C (typical
4
)
Bandwidth limit
Select from OFF, Auto, 4 KHz, 400 Hz, and 40 Hz
Cutoff characteristics:
–12 dB/OCT (typical
4
, excluding AUTO)
Cutoff frequency (fc) when set to AUTO
Sample rate of 100 kHz or higher:
fc = 40 kHz
Sample rate of 100 Hz to 50 kHz:
fc = 40% of the sampling rate
Sample rate of 50 Hz or less:
fc = 20 Hz
Cutoff characteristics when set to AUTO:
–65 at 2
×
fc (typical
4
)
Probe attenuation setting
Voltage probe
1:1, 10:1, 100:1, or 1000:1
Current probe
10 A:1 V (for the 700937/701933), 100 A:1 V (for the 701930/701931)
Compatible probes/cables Connection cable (for low voltage 1:1)
366926 (non-isolated type BNC-alligator clip
×
2: For measuring low voltage less than or equal to
42 Vpeak)
Voltage probe (10:1 passive probe) 701940
17 to 46 pF: For measuring 600 V (DC+ACpeak) or less
Current probe (power can be supplied from the DL750/DL750P)
700937 (15 A), 701930 (150 A), 701931 (500 A), 701933 (30 A)
Sensor supply current
OFF or 4 mA
±
10% (approx. 22 VDC)
(voltage)
Applicable acceleration
Built-in amplifier type
sensor
Kistler Instrument Corporation: Piezotron, PCB Piezotronics Incorporated: ICP, ENDEVCO
Corporation: ISOTRON, etc.
19.12 Module Specifications