11-4
IM 701210-06E
Notes When Recalling Data Using the History Memory Function
• You cannot use the history memory function while waveform acquisition is in
progress.
• You cannot restart the waveform acquisition when the history memory menu is
displayed.
• Settings are restricted by the following condition: End Record
≤
Select Record
≤
Start
Record.
• If you load a waveform record from the external storage medium, the loaded
waveform becomes Record 0. In the case of multiple records (sequential store), the
records will be loaded sequentially, with the newest record as 0.
• Computation and automated measurement of waveform parameters are performed on
the record No. specified by Select Record. Analysis of old data is possible as long as
the history memory content remains unchanged after acquisition is restarted.
However, for average display, computation is not performed again when you specify a
different record.
• The time displayed in Show Map is the waveform trigger time. The information varies
depending on the trigger mode as follows:
Trigger Mode
Condition
Time Displayed in Show Map
Auto/Auto Level
Roll mode
Stop time
Single
Roll mode, trigger not activated
Stop time
Log
Start time
• When displaying all waveforms, if the selected number of records is large, it may take
an extended period for the display to complete. In this case,
is indicated at the
upper left corner of the screen. To abort the operation, set Display Mode to One.
• Turning OFF the power clears the contents of the history memory. However, if the
acquisition memory backup is enabled, the contents of the history memory are backed
up.
For details on the acquisition memory backup function, see section 7.10.
11.1 Displaying History Waveforms