19-39
IM 701210-06E
Specifications
19
Specifications by Measurement Modes
Item
Specifications
Frequency
Measurable frequency range
0.01 Hz to 200 kHz
Selectable vertical axis sensitivity range
0.1 Hz/div to 50 kHz/div (1-2-5 steps)
Minimum resolution
0.00 1Hz
RPMs
Measurable RPMs range
0.01 rpm to 100000 rpm (where the input frequency is DC to 200 kHz).
Selectable vertical axis sensitivity range
0.1 rpm/div to 10 krpm/div (1-2-5 steps)
Computing method
Computed from the frequency based on the number of pulses per rotation
RPMs = Frequency/(pulse/rotate value)
×
60
Selectable pulse/rotate range
1 to 99999
RPSs
Measurable RPSs range
0.001 rps to 2000 rps (where the input frequency is DC to 200 kHz).
Selectable vertical axis sensitivity range
0.01 rps/div to 200 rps/div (1-2-5 steps)
Computing method
Computed from the frequency based on the number of pulses per rotation
RPSs = Frequency/(pulse/rotate value)
Selectable pulse/rotate range
1 to 99999
Period
Measurable period range
5 ms to 50 s (where the minimum pulse width is 2
µ
s)
Selectable vertical axis sensitivity range
10
µ
s to 500
µ
s, 1 ms to 500 ms, 1 to 5 s/div (1-2-5 steps)
Minimum resolution
0.1
µ
s
Duty cycle
Measurable duty cycle range
0 to 100%
Selectable vertical axis sensitivity range
1 to 20% (1-2-5 steps)
Measurable frequency range
0.1 Hz to 200 kHz
Measurement pulse selection
Select positive or negative pulse
Minimum resolution
0.01%
Power supply frequency
Measurable frequency range
30 Hz to 70 Hz (when the center frequency is 50 Hz), 40 Hz to 80 Hz (when the
center frequency is 60 Hz), 380 Hz to 420 Hz (when the center frequency is 400
Hz)
Selectable vertical axis sensitivity range
0.1 Hz/div to 2 Hz/div (0.01 Hz resolution)
Center frequency setting
Select 50 Hz, 60 Hz, or 400 Hz
Minimum resolution
0.01 Hz
Pulse width
Measurable pulse width
2
µ
s to 50 s (where the input frequency is up to 200 kHz)
Selectable vertical axis sensitivity range
10
µ
s to 500
µ
s, 1 ms to 500 ms, 1 to 5 s/div (1-2-5 steps)
Measurement pulse selection
Select positive or negative pulse
Minimum resolution
0.1
µ
s
Pulse integration
Maximum pulse count
2
×
10
9
pulses
Selectable vertical axis sensitivity range
500.0E+18 value/div to 10.00E–21 value/div (1-2-5 range: total of 123 ranges)
Frequency measuring range
0.1 Hz to 200 kHz (where the minimum pulse width is 2
µ
s)
Computation function
Set the physical amount per pulse and display by converting the values into
physical values such as distance and flow rate.
Selectable Unit/Pulse range
–9.9999E+30 to +9.9999E+30
Counter reset
Manual reset and over-limit reset
Velocity
Selectable vertical axis sensitivity range
500.0E+18 value/div to 10.00E–21 value/div (1-2-5 range: total of 123 ranges)
Computing method
Set the amount of displacement per pulse and compute the velocity from the
frequency
Automatic unit time conversion of s, min, and hour.
Selectable Distance/Pulse range
–9.9999E+30 to +9.9999E+30
19.12 Module Specifications