![YOKOGAWA ScopeCorder DL750 User Manual Download Page 324](http://html.mh-extra.com/html/yokogawa/scopecorder-dl750/scopecorder-dl750_user-manual_911347324.webp)
19-36
IM 701210-06E
Item
Specifications
Table of Cutoff Frequency Characteristics of the Anti-Aliasing Filter (AAF)
When the filter is set to Auto, the anti-aliasing filter and low-pass filter are automatically set
according to the sample rate.
Sample Rate AAF
Low-pass filter
100 kS/s
40 kHz
OFF
50 kS/s
20 kHz
OFF
20 kS/s
8 kHz
OFF
10 kS/s
4 kHz
4 kHz
5 kS/s
2 kHz
4 kHz
2 kS/s
800 Hz
4 kHz
1 kS/s
400 Hz
400 Hz
500 S/s
200 Hz
400 Hz
200 S/s
80 Hz
400 Hz
100 S/s
40 Hz
40 Hz
50 S/s
20 Hz
40 Hz
20 S/s to 5 S/s
20 Hz
40 Hz
2 S/s or less
20 Hz
40 Hz
Ext sample
40 kHz
OFF
1
Value measured under standard operating conditions (section 19.11).
H
L
2.
3.
BNC
Combined with the 10:1 passive probe
(701940)
Direct input (cable not complying with
the safety standard)
H
L
2.
3.
4
Typical value represents a typical or average value. It is not strictly warranted.
WARNING
• Do not apply input voltage exceeding the maximum input voltage, withstand
voltage, or allowable surge voltage.
• To prevent the possibility of electric shock, be sure to furnish protective earth
grounding of the DL750/DL750P.
• To prevent the possibility of electric shock, be sure to fasten the module screws.
Otherwise, the electrical and mechanical protection functions will not be
activated.
• Avoid continuous connection under an environment in which the allowable surge
voltage may occur.
19.12 Module Specifications