19-25
IM 701210-06E
Specifications
19
Item
Specifications
Compatible probes/cables Connection cable (for high voltage 1:1): Recommended 1
701901 (isolated type BNC-safety alligator clip adapter
×
2: For measuring 850 V (DC+ACpeak)
or less),
701954 (alligator clip (dolphin type) red/black 2-piece set) is required separately
Voltage probe (10:1 safety probe): Recommended 2
700929 (10:1 safety probe) .20 to 45 pF: For measuring 1000 V (DC+ACpeak) or less
Current probe (power can be supplied from the DL750/DL750P)
700937 (15 A), 701930 (150 A), 701931 (500 A), 701933 (30 A)
1.
Value measured under standard operating conditions (section 19.11).
H
L
Combined with the 700929
2.
3.
4.
700929
H
L
701901
701954
5.
6.
7.
Combined with the 701954
H
L
Direct input (cable not complying with the safety standard)
8.
9.
BNC
Withstand voltage: 3700 Vrms for 1 minute
10. Typical value represents a typical or average value. It is not strictly warranted.
WARNING
• When applying high voltage using this module, use the 1:1 safety cable
(combination of 701901 and 701954) or the isolated probe (700929).
• The Measurement Category of the direct input of this module is 400 Vrms-CAT
II on the low side and 700 Vrms-CAT II on the high side. Use caution because
the overvoltage category differs between the low and high sides.
• Do not apply input voltage exceeding the maximum input voltage, withstand
voltage, or allowable surge voltage.
• To prevent the possibility of electric shock, be sure to furnish protective earth
grounding of the DL750/DL750P.
• To prevent the possibility of electric shock, be sure to fasten the module screws.
Otherwise, the electrical and mechanical protection functions will not be
activated.
• Avoid continuous connection under an environment in which the allowable surge
voltage or higher voltage may occur.
19.12 Module Specifications