11-38
IM 701210-06E
11.7 Performing Statistical Processing
<For a description of this function, refer to page 2-57.>
Procedure
MEASURE CURSOR
RESET
SELECT
CH
ALL CH
MODE
POSITION
SIMPLE/ENHANCED
ACQ
START/STOP
SETUP
DISPLAY
ZOOM
DUAL
CAPTURE
HISTORY
MATH
1
CH
2
CH
3
CH
4
CH
5
CH
6
CH
7
CH
8
CH
9
CH
10
CH
11
CH
12
CH
13
CH
14
CH
15
CH
16
DELAY
MANUAL TRIG
V/DIV
TIME/DIV
TRIGGER
TRIG D
VERTICAL
HORIZONTAL
CAL
X-Y
SEARCH
7
DSP 1
8
9
DSP 2
6
5
4
1
0
2
3
ENTER
m
DSP 6
DSP 5
LOGIC A
LOGIC B
EVENT
DSP 3
DSP 4
EXP
GO/NO-GO
ACTION
FILE
SHIFT
Setting Normal Statistical Processing
1.
Press
MEASURE
.
2.
Press the
Mode
soft key to select
Statistics
.
3.
Press the
Item Setup
soft key. The measurement parameter dialog box and the
measurement channel menu appear.
4.
Press the
Trace
soft key and use the
jog shuttle
and
SELECT
to select the
measurement channel.
5.
Turn the
jog shuttle
to move the cursor to the parameter you wish to turn ON.
6.
Press
SELECT
to turn on the parameter selected in step 5.
You can turn OFF all parameters at once by pressing the
All Clear
. You can
copy the current parameter settings to all traces by selecting “
Copy to All
Trace
.”
7.
Repeat steps 4 to 6 as many times as necessary. Press
ESC
.
The setting of other parameters is the same as the automated measurement of
waveform parameters (see step 8 and following steps in section 11.6).
DSP1 to DSP6 are optional.