![YOKOGAWA ScopeCorder DL750 User Manual Download Page 73](http://html.mh-extra.com/html/yokogawa/scopecorder-dl750/scopecorder-dl750_user-manual_911347073.webp)
11-35
IM 701210-06E
Waveform Analysis/Search
11
Automated measurement of various waveform parameters are performed on the data
stored to the acquisition memory. Up to 48000 data points among the parameters that
are automatically measured can also be saved to a file. (For details, see section 13.10.)
Burst1
Width
Distal Line (90%)
Proximal Line (10%)
Measurement Range
Set the measurement range
(T-Range) to match the burst
width of the signal being measured.
Burst1, Burst2: Burst Width
(Bst1, Bst2)
T1 cursor
T2 cursor
Mesial Line (50%)
Burst2
Width
• Other Measurement Parameters
Int1TY:
The area under the positive amplitude
Int2TY:
The area under the positive amplitude – the area under the negative
amplitude
Int1XY:
The summation of the triangular area of the X-Y waveform
Int2XY:
The summation of the trapezoidal area of the X-Y waveform
For details regarding the area calculations, see Appendix 3, “How to Calculate the Area of a
Waveform.”
Setting the Delay
The time difference between the rising or falling edges of trace waveforms or trigger
points is called the delay between channels.
Reference
Waveform
Measurement
Waveform
1 count
N1 count
1 count
N2 count
Delay between Channels
• Edge Polarity
• Edge Count
Setting Example of Reference
Waveform
Setting Example of Measurement
Waveform
Measurement Range
: (Rising)
: N1(1 to 9)
• Edge Polarity
• Edge Count
: (Falling)
: N2(1 to 9)
Mesial Line
• If Mode is set to Time or Degree, measurements will be made. The default setting is
OFF.
Time:
Display the delay between channels as a time
Degree: Display the delay between channels as an angle
Degree = Delay (s)/Period (s)
×
360 (deg). The period is that of a reference
waveform.
• Select (rising edge) or (falling) for the slope of the edge to be detected using
Edge Polarity. The default setting is rising.
• Set the number of edges to detect before actually considering it a detection point in
Edge Count. The range is an integer from 1 to 9. The default value is 1.
11.6 Automated Measurement of Waveform Parameters