18-3
IM 701210-06E
Troubleshooting, Maintenance, and Inspection
18
18.2 Messages and Corrective Actions
Code
Message and Corrective Action
Section
93
Realtime recording is being prepared.
7.7
94
Media information is being read.
13.5
95
Failed to retrieve the data. Execute again.
Chapter 11
96
This setting not possible on DSP channels.
Chapter 15
97
The maximum sampling rate when using DSP channels is 5 MS/s.
Chapter 15
Slower sample rate than 2MS/s when set a DSP display.
279
This command cannot be used in the current mode.
Chapter 9
Errors in Execution (700 to 799)
Code
Message and Corrective Action
Section
701
Long copy is not possible when waveform acquisition is in progress.
7.1, 12.2
Press the START/STOP key to stop the waveform acquisition first.
703
Undo is not possible since data that existed immediately before initialization and auto setup is not
4.4, 4.5
available.
704
Can not be executed while running. Press START/STOP key to stop acquisition.
7.1
707
Can not Start while data out. Wait until output is completed.
7.1
708
Can not data out while running. Press START/STOP key to stop acquisition.
7.1
709
Can not detect listener. Check GP-IB connector.
–
710
File not found. Check the file.
Chapter 13
711
Cannot manipulate files while image printing is in progress. Wait until image printing is complete.
Chapter 13
712
Can not compress this hardcopy image. Turn off compression switch.
13.11
713
Calibration failure. Disconnect the input and execute again.
4.6
If it fails again, servicing is necessary.
714
Temperature module calibration failed. Calibration value is invalid. Check the input signal.
4.6
716
Realtime recording to the internal hard disk is valid when the sampling rate is slower than the values
7.7
shown below.
1~2 CH : 100 kS/s, 3CH : 50 kS/s,
4~5CH : 20 kS/s, 6~11CH : 10 kS/s,
12~18CH : 5 kS/s
717
Too many channels for the current T/div setting to realtime record.
7.7
Decrease the number of channels by turning them OFF.
718
Too many channels to realtime record at the current record length.
7.7
Turn OFF channels to reduce the number of channels or shorten the record length.
721
Can not operate while data out. Wait until output is completed.
7.7
723
Cannot load the realtime record.
7.7,13.14
724
Balancing failed.
–
* indicates the channel number of the channel for which automatic balancing failed.
725
Cannot be executed. Acquisition not active.
7.1
727
Insufficient output data. Increase Mag or widen the Time Range interval.
12.2
728
Image is being printed. Abort or wait until printing is complete.
Chapter 13
730
Pattern is not specified.
11.4
732
Cannot be executed while computation is in progress.
Chapter 10
733
Failed to measure statistics.
11.7
Waveform data may be missing.
If Cycle Statistics is specified, the instrument may be configured in a way that fails to detect the cycle.
735
Executing file Load, Save, or Format. Abort or wait until it is complete.
Chapter 13
736
Image is being printed or saved. Abort the operation or wait until the execution of the command is
Chapter 13
complete.
737
A setup of an action trigger can’t be carried out under the state of on.
6.18
738
Cannot be executed when the dual capture setting is ON.
7.6
739
Cannot be executed when realtime record or print setting is ON.
7.7
740
A time base can’t be carried out under the state of the outside clock.
5.14
741
A start by the START key can’t be done in the GO-NOGO mode.
11.8, 11.9