10-6
IM 701210-06E
Explanation
This function converts CH1 to CH16 or Math1 to Math7 waveform to a digital signal (1s
and 0s) according to the specified threshold level.
1
0
Upper
Lower
Setting the Computation Range: Start Point/End Point
By default, the measurement range is
±
5 divisions of the display frame on the time axis.
You can limit this range.
The concept of the computation range is analogous to the concept of the selectable
range of cursor display position in cursor measurement.
For details, see section 11.5, “Selectable Range of Cursor Position.”
Setting the Unit
Unit can be set arbitrarily using up to four characters. The specified characters are
applied to the scale values.
Linear Scaling
When performing computation on a channel that has linear scaling set, the computation
is performed on the scaled value.
Notes When Performing Computation
Computation is not performed again when you change Start Point or End Point while
computation is stopped. Be sure to press the Exec soft key to perform the computation
again. Otherwise, the waveform will not be displayed correctly when the screen is
redrawn.
Note
While computation is in execution,
is displayed at the upper left corner of the screen.
10.2 Binary Computation