11-51
IM 701210-06E
Waveform Analysis/Search
11
• Setting the Number of Actions
13.
Use the
jog shuttle
and
SELECT
to set Sequence to Single or Continue.
• Setting the Number of Determinations
14.
Use the
jog shuttle
and
SELECT
to set ACQ Count.
• Enabling/Disabling the External Input Signal
15.
Use the
jog shuttle
and
SELECT
to set Remote to OFF or ON.
Selecting the Action
16.
Press the
Action
soft key.
17.
Use the
jog shuttle
and
SELECT
to set each action ON/OFF.
18.
If you turn ON Save to File in step 17, use the
jog shuttle
and
SELECT
to set
the file save format to Binary, ASCII, or Float.
Setting the Determination Range
19.
Press the
Time Range1/Time Range2
soft key to set the jog shuttle control to
Time Range1.
20.
Turn the
jog shuttle
to set the determination start point.
21.
Likewise, set Time Range2 (determination end point).
Executing the GO/NO-GO Determination
22.
Press the
Exec
soft key to execute the GO/NO-GO determination. The soft key
changes to
Abort
. Waveform acquisition stops automatically, when
determination is finished,
To stop the GO/NO-GO determination, press the
Abort
soft key or
START/
STOP
.
In addition, if the following action is specified, the following icon is displayed at
the upper left corner of the screen when an action is activated.
Image/Save to File:
Save destination medium (
FD,
Zip disk,
PC
card,
external SCSI device,
internal HD,
network drive, or
USB storage device)
Send Mail:
11.9 GO/NO-GO Determination Using Measured Waveform Parameters