11-20
IM 701210-06E
For Angle (Degree) Cursors (When Not Displaying the X-Y Waveform)
MEASURE CURSOR
RESET
SELECT
CH
ALL CH
MODE
POSITION
SIMPLE/ENHANCED
ACQ
START/STOP
SETUP
DISPLAY
ZOOM
DUAL
CAPTURE
HISTORY
MATH
1
CH
2
CH
3
CH
4
CH
5
CH
6
CH
7
CH
8
CH
9
CH
10
CH
11
CH
12
CH
13
CH
14
CH
15
CH
16
DELAY
MANUAL TRIG
V/DIV
TIME/DIV
TRIGGER
TRIG D
VERTICAL
HORIZONTAL
CAL
X-Y
SEARCH
7
DSP 1
8
9
DSP 2
6
5
4
1
0
2
3
ENTER
m
DSP 6
DSP 5
LOGIC A
LOGIC B
EVENT
DSP 3
DSP 4
EXP
GO/NO-GO
ACTION
FILE
SHIFT
1.
Press
CURSOR
.
Selecting the Cursor Type
2.
Press the
Type
soft key. The cursor type selection menu appears.
3.
Press the
Degree
soft key.
Selecting the Waveform to Be Measured
4.
Press the
Trace
soft key. The trace selection menu appears.
5.
Press the soft key corresponding to the desired waveform to select the
waveform to be measured.
Moving the Cursor
6.
Press the
Cursor1/Cursor2
soft key to set the jog shuttle control to Cursor1.
7.
Turn the
jog shuttle
to move Cursor1.
8.
Likewise, move Cursor2. If the jog shuttle control is set to both Cursor1 and
Cursor2, both cursors are moved.
Moving the Reference Cursor
9.
Press the
Ref1/Ref2
soft key to set the jog shuttle control to Ref1.
10.
Turn the
jog shuttle
to move Ref1.
11.
Likewise, move Ref2. If the jog shuttle control is set to both Ref1 and Ref2, both
cursors are moved.
Setting the Reference Angle
12.
Press the
Ref Value
soft key.
13.
Turn the
jog shuttle
to set the reference angle in the range of 1 to 720.
11.5 Measuring Waveforms Using Cursors