Powers up the device
•
Opens an onscreen Power-down dialog with options to Power down, Sleep, or
Cancel (brief press)
•
Sets the device into Display power-saving (Sleep) mode (brief press followed by 30
seconds of inactivity)
Note
The device will not enter the Sleep (Display power-saving) mode while an
alarm condition is active or when NIBP intervals or continuous
measurements are in progress. The device also exits this mode if an alarm
occurs.
The LED in the center of the power plug symbol indicates the battery charging status:
•
Green indicates that AC power is present and that the battery is fully charged.
•
Amber indicates that AC power is present and that the battery is charging.
Power up the monitor
The monitor runs a brief diagnostic self-test each time it powers up.
WARNING
Equipment failure risk. The monitor includes a fan that
circulates air through the device. If the fan does not run when you power
up the device, remove it from use and inform qualified service personnel
immediately. Do not use the monitor until the problem is corrected.
WARNING
To ensure patient safety, listen for an audible indicator and
watch for visual alerts at power-up at least once daily. Correct any system
errors before using the monitor. In addition to the audible indicator, the
monitor LED light bar illuminates to alert you of alarms. Amber indicates a
low-level alarm. Flashing amber indicates a medium-level alarm. Flashing
red indicates a high-level alarm.
WARNING
Always observe the monitor during power-up. If any display
fails to illuminate properly, or if an error code displays, inform qualified
service personnel immediately, or call your nearest Welch Allyn Customer
Service or Technical Support facility. Do not use the monitor until the
problem is corrected.
CAUTION
Always use the monitor with an adequately charged and
properly functioning battery.
CAUTION
During continuous monitoring, keep the monitor connected to
AC power at all times.
CAUTION
Use only a Class I (grounded) AC power cord for powering this
monitor.
1. Press to power up the monitor.
Following a successful self-test, the monitor displays the Welch Allyn logo, the LED
light bar (located on the handle) flashes, and a power-up tone sounds. The startup
screen then appears with the following banner across the bottom.
56 Startup
Welch Allyn Connex
®
Devices
Summary of Contents for Connex IWS
Page 1: ...Welch Allyn Connex Devices Directions for use Software version 2 3X...
Page 10: ...4 Introduction Welch Allyn Connex Devices...
Page 14: ...8 Symbols Welch Allyn Connex Devices...
Page 20: ...14 Screen elements Welch Allyn Connex Devices...
Page 28: ...22 About warnings and cautions Welch Allyn Connex Devices...
Page 72: ...66 Navigation Welch Allyn Connex Devices...
Page 96: ...90 Profiles Welch Allyn Connex Devices...
Page 106: ...100 Patient data management Welch Allyn Connex Devices...
Page 120: ...114 Alarms Welch Allyn Connex Devices...
Page 184: ...178 Patient monitoring Welch Allyn Connex Devices...
Page 212: ...206 Advanced settings Welch Allyn Connex Devices...
Page 244: ...238 Specifications Welch Allyn Connex Devices...
Page 250: ...244 Standards and compliance Welch Allyn Connex Devices...
Page 252: ...246 Guidance and manufacturer s declaration Welch Allyn Connex Devices...
Page 266: ......