PRO 6000 model
PRO 4000 model
•
Blinking green: The battery pack is fully charged.
•
Steady green: The battery pack is ready to charge, but
dock is empty or thermometer is improperly docked.
•
No LED/Off: Not ready to charge. The monitor is not
connected to AC power and is not powered on, or the
monitor has disabled charging.
•
Amber: The battery pack is charging.
•
No LED/Off: Not ready to charge. The monitor is not
connected to AC power and is not powered on, or the
monitor has disabled charging.
Note
The battery pack continues to charge while the monitor is in Display power
saving mode.
Note
It is strongly recommended that you use only the Welch Allyn rechargeable
battery pack in the thermometer because the dock cannot charge other
batteries.
Temperature mode selection
The monitor with the temperature module takes a patient temperature in either
Predictive (Normal) or Direct mode. The default setting is the Predictive mode.
WARNING
Inaccurate measurement risk. To ensure optimal accuracy,
always confirm that the correct mode and site are selected.
WARNING
Patient injury risk. Do not exceed the recommended
temperature measurement durations in Direct mode. Continuous
measurement durations of 3 minutes at the oral and rectal sites and 5
minutes at the axillary site are recommended for accurate measurement.
Do not continuously measure beyond 10 minutes in any mode.
Predictive mode
Is a one-time measurement that takes a temperature in approximately 6 to 15 seconds.
Removing the probe from the probe well, loading a probe cover, and holding the probe
tip in place at the measurement site initiates a Predictive mode measurement. The
monitor sounds a tone to indicate the end of a predictive measurement.
Direct mode
Provides continual temperature measurements. For oral and rectal measurements, it is
recommended to measure temperature until the temperature stabilizes or for 3 minutes.
For axillary measurements, it is recommended to measure temperature until the
temperature stabilizes or for 5 minutes. The monitor changes to Direct mode
approximately 60 seconds after the probe is removed from the probe well.
Note
The monitor does not retain Direct mode temperatures in
memory.Therefore, it is important to note the temperature before removing
the probe from the measurement site and then manually record it in the
patient record.
After 10 minutes of using the Direct mode, the monitor generates a technical alarm
condition and clears the measurement.
162 Patient monitoring
Welch Allyn Connex
®
Devices
Summary of Contents for Connex IWS
Page 1: ...Welch Allyn Connex Devices Directions for use Software version 2 3X...
Page 10: ...4 Introduction Welch Allyn Connex Devices...
Page 14: ...8 Symbols Welch Allyn Connex Devices...
Page 20: ...14 Screen elements Welch Allyn Connex Devices...
Page 28: ...22 About warnings and cautions Welch Allyn Connex Devices...
Page 72: ...66 Navigation Welch Allyn Connex Devices...
Page 96: ...90 Profiles Welch Allyn Connex Devices...
Page 106: ...100 Patient data management Welch Allyn Connex Devices...
Page 120: ...114 Alarms Welch Allyn Connex Devices...
Page 184: ...178 Patient monitoring Welch Allyn Connex Devices...
Page 212: ...206 Advanced settings Welch Allyn Connex Devices...
Page 244: ...238 Specifications Welch Allyn Connex Devices...
Page 250: ...244 Standards and compliance Welch Allyn Connex Devices...
Page 252: ...246 Guidance and manufacturer s declaration Welch Allyn Connex Devices...
Page 266: ......