Nellcor specifications (for additional information, refer to the directions for use that
accompanies each accessory)
Low Perfusion: 0.02% to 20% ± 2 digits
Saturation accuracy (sensors)
Sensor
Accuracy
60% to 80%
MAX-AI, MAX-
PI, MAX-II
±3 digits
70% to 100%
DS-100A
±3 digits
D-YS
Infants, Pediatrics, Adults: ±3 digits
Neonates: ±4 digits
D-YSE
±4 digits
D-YSPD
±4 digits
MAX-AI, MAX-
PI, MAX-II
±2 digits
OXI-A/N
Adults: ±3 digits
Neonates: ±4 digits
OXI-P/I
±3 digits
Electrical/Optical Specifications
Nellcor pulse oximetry sensors contain light-emitting diodes
(LEDs) that emit red light at a wavelength of approximately
660 nm, and infrared light at a wavelength of approximately
900 nm. The total optical output power of the sensor LEDs is
less than 15 mW. This information might be useful to
clinicians, such as those performing photodynamic therapy.
Pulse rate
Unit of measure
beats per minute
Measurement range
20 to 250 beats per minute
Accuracy
±3 digits
Masimo specifications (for additional information, refer to the directions for use that
accompanies each accessory)
SpO2
Note
Contact the sensor manufacturer for
further SpO2 clinical testing information.
Note
Refer to sensor manufacturers’ directions
for use for further accuracy information.
Unit of measure
%
Measurement range
0% to 100%
Directions for use
Specifications 227
Summary of Contents for Connex IWS
Page 1: ...Welch Allyn Connex Devices Directions for use Software version 2 3X...
Page 10: ...4 Introduction Welch Allyn Connex Devices...
Page 14: ...8 Symbols Welch Allyn Connex Devices...
Page 20: ...14 Screen elements Welch Allyn Connex Devices...
Page 28: ...22 About warnings and cautions Welch Allyn Connex Devices...
Page 72: ...66 Navigation Welch Allyn Connex Devices...
Page 96: ...90 Profiles Welch Allyn Connex Devices...
Page 106: ...100 Patient data management Welch Allyn Connex Devices...
Page 120: ...114 Alarms Welch Allyn Connex Devices...
Page 184: ...178 Patient monitoring Welch Allyn Connex Devices...
Page 212: ...206 Advanced settings Welch Allyn Connex Devices...
Page 244: ...238 Specifications Welch Allyn Connex Devices...
Page 250: ...244 Standards and compliance Welch Allyn Connex Devices...
Page 252: ...246 Guidance and manufacturer s declaration Welch Allyn Connex Devices...
Page 266: ......