3. Touch the
IPI
vertical tab.
4. Ensure that the Limits control is set to ON.
Note
If any parameter's alarm limit control is set to OFF, you
cannot adjust alarm limits on the Alarm tab, and no visual or
audio signals will occur for that specific parameter.
5. Adjust the IPI limits setting as desired. Enter the desired lower alarm limit for IPI
using the up/down arrow keys or the keypad.
6. Touch the
Home
tab.
The new alarm settings take effect immediately.
Acoustic respiration rate (RRa)
Monitors configured with Masimo rainbow Acoustic Monitoring technology continuously
and noninvasively measure respiration rate based on respiratory sounds produced in the
upper airway. Respiratory sounds are detected through a sensor applied to the patient's
neck.
Note
RRa is available only in the Continuous Monitoring profile.
Note
Masimo SpO2 monitoring is required when monitoring RRa.
Note
RRa monitoring is only for adult patients who weigh more than 66 pounds
(30 kilograms).
Note
Monitors configured with RRa cannot be configured with CO2 or SpHb.
RRa frame
The RRa frame displays the following:
•
Respiration rate in breaths per minute (BPM)
•
Respiration Indicator
•
Alarm limit control
•
Averaging control toggle providing the options of slow, medium, fast, trending, and
off
Note
RRa is available only when the Continuous Monitoring profile and Adult
patient type are selected.
Respiration Indicator
The Respiration Indicator displays the level of sound detected by the RRa sensor. A
higher level of sound causes more bars to illuminate.
Directions for use
Patient monitoring 131
Summary of Contents for Connex IWS
Page 1: ...Welch Allyn Connex Devices Directions for use Software version 2 3X...
Page 10: ...4 Introduction Welch Allyn Connex Devices...
Page 14: ...8 Symbols Welch Allyn Connex Devices...
Page 20: ...14 Screen elements Welch Allyn Connex Devices...
Page 28: ...22 About warnings and cautions Welch Allyn Connex Devices...
Page 72: ...66 Navigation Welch Allyn Connex Devices...
Page 96: ...90 Profiles Welch Allyn Connex Devices...
Page 106: ...100 Patient data management Welch Allyn Connex Devices...
Page 120: ...114 Alarms Welch Allyn Connex Devices...
Page 184: ...178 Patient monitoring Welch Allyn Connex Devices...
Page 212: ...206 Advanced settings Welch Allyn Connex Devices...
Page 244: ...238 Specifications Welch Allyn Connex Devices...
Page 250: ...244 Standards and compliance Welch Allyn Connex Devices...
Page 252: ...246 Guidance and manufacturer s declaration Welch Allyn Connex Devices...
Page 266: ......