NIBP specifications
Pulse rate accuracy (using blood pressure
determination)
±5.0% (±3 beats per minute)
Overpressure cutoff
Adult: 300 mmHg ±15 mmHg (40.0 kPa ±2.0 kPa)
Pediatric: 300 mmHg ±15 mmHg (40.0 kPa ±2.0 kPa)
Neonate: 150 mmHg (20.0 kPa) maximum
Note
NIBP performance can be affected by extremes of humidity, temperature, and altitude.
SureTemp Plus temperature module specifications
Units of measure
°F, °C; user-selectable
Temperature range
80° F to 110° F (26.7° C to 43.3° C)
Temperature measurement range
Accuracy
Less than 37.0 °C
±0.2 °C
37.0 °C to 39.0 °C
±0.1 °C
Greater than 39.0 °C
±0.2 °C
Less than 96.4 °F
±0.4 °F
96.4 °F to less than 98.0 °F
±0.3 °F
98.0 °F to 102.0 °F
±0.2 °F
102.0 °F to 106.0 °F
±0.3 °F
Greater than 106.0 °F
±0.4 °F
Braun ThermoScan PRO thermometer specifications (refer to manufacturer's directions for use
for additional information)
Units of measure
°F, °C; user-selectable
Temperature range
68° F to 108° F (20° C to 42.2° C)
Calibration accuracy
•
±0.4° F (±0.2° C) for temperatures ranging from 95.9° F to 107.6° F
(35.5° C to 42° C)
•
±0.5° F (±0.3° C) for temperatures outside of this range
Display resolution
0.1° F or 0.1° C
Directions for use
Specifications 225
Summary of Contents for Connex IWS
Page 1: ...Welch Allyn Connex Devices Directions for use Software version 2 3X...
Page 10: ...4 Introduction Welch Allyn Connex Devices...
Page 14: ...8 Symbols Welch Allyn Connex Devices...
Page 20: ...14 Screen elements Welch Allyn Connex Devices...
Page 28: ...22 About warnings and cautions Welch Allyn Connex Devices...
Page 72: ...66 Navigation Welch Allyn Connex Devices...
Page 96: ...90 Profiles Welch Allyn Connex Devices...
Page 106: ...100 Patient data management Welch Allyn Connex Devices...
Page 120: ...114 Alarms Welch Allyn Connex Devices...
Page 184: ...178 Patient monitoring Welch Allyn Connex Devices...
Page 212: ...206 Advanced settings Welch Allyn Connex Devices...
Page 244: ...238 Specifications Welch Allyn Connex Devices...
Page 250: ...244 Standards and compliance Welch Allyn Connex Devices...
Page 252: ...246 Guidance and manufacturer s declaration Welch Allyn Connex Devices...
Page 266: ......