Message
Possible cause
Suggested action
The bed sensor is upside down.
The bed sensor was not placed properly under
the mattress, mattress cover, or mattress pad
Turn the sensor right-side up.
Cannot measure patient. Lost or
unstable signal.
The sensor is not under the patient's chest
(Low confidence condition)
Reposition the sensor under the
patient's chest.
The sensor is turned 90 degrees, placed
vertically (Low confidence condition)
Place the sensor horizontally
under the patient's mattress,
with the cable extending toward
the head of the bed.
There is excessive patient movement (Low
confidence condition)
Check the patient.
Replace the bed sensor.
An unqualified mattress type is in use
Switch to a qualified mattress
type.
Cannot measure respiration.
The sensor is not under the patient's chest
(Low confidence condition)
Adjust the position of the sensor
under the patient's chest.
The sensor is turned 90 degrees, placed
vertically (Low confidence condition)
Place the sensor horizontally
under the patient's mattress,
with the cable extending toward
the head of the bed.
There is excessive patient movement (Low
confidence condition)
Check the patient.
Replace the bed sensor.
Use an alternative sensor to
monitor respiration.
An unqualified mattress type is in use
Switch to a qualified mattress
type.
Cannot measure pulse rate.
The sensor is not under the patient's chest
(Low confidence condition)
Adjust the position of the sensor
under the patient's chest.
The sensor is turned 90 degrees, placed
vertically (Low confidence condition)
Place the sensor horizontally
under the patient's mattress,
with the cable extending toward
the head of the bed.
There is excessive patient movement (Low
confidence condition)
Check the patient.
Replace the bed sensor.
Use an alternative sensor to
monitor pulse rate.
An unqualified mattress type is in use
Switch to a qualified mattress
type.
EarlySense not functional.
A module error occurred
Call for service.
Bed sensor expires in ...
The bed sensor expires soon
Replace the bed sensor before it
expires.
The sensor has expired.
The bed sensor has expired
Replace the sensor.
208 Troubleshooting
Welch Allyn Connex
®
Devices
Summary of Contents for Connex IWS
Page 1: ...Welch Allyn Connex Devices Directions for use Software version 2 3X...
Page 10: ...4 Introduction Welch Allyn Connex Devices...
Page 14: ...8 Symbols Welch Allyn Connex Devices...
Page 20: ...14 Screen elements Welch Allyn Connex Devices...
Page 28: ...22 About warnings and cautions Welch Allyn Connex Devices...
Page 72: ...66 Navigation Welch Allyn Connex Devices...
Page 96: ...90 Profiles Welch Allyn Connex Devices...
Page 106: ...100 Patient data management Welch Allyn Connex Devices...
Page 120: ...114 Alarms Welch Allyn Connex Devices...
Page 184: ...178 Patient monitoring Welch Allyn Connex Devices...
Page 212: ...206 Advanced settings Welch Allyn Connex Devices...
Page 244: ...238 Specifications Welch Allyn Connex Devices...
Page 250: ...244 Standards and compliance Welch Allyn Connex Devices...
Page 252: ...246 Guidance and manufacturer s declaration Welch Allyn Connex Devices...
Page 266: ......